Brief Procedures
TDS 410, 420 & TDS 460 Service Manual (Company Confidential Final Draft)
4Ć5
First, display the CAL menu.
Second, verify Pass
status for the
adjustment
sections.
Third, run a signal
path compensation
and verify status is
Pass.
Figure 4-2: Verifying Adjustments and Signal Path Compensation
This oscilloscope lets you compensate the internal signal path used to acquire
the waveforms you acquire and measure. By executing the signal path com-
pensation feature (SPC), you optimize the oscilloscope capability to make
accurate measurements based on the ambient temperature.
You
must run an SPC anytime you wish to ensure that the measurements you
make are made with the most accuracy possible. You should also run an SPC
if the temperature has changed more than 5
_
C since the last SPC was per-
formed.
f.
Run the signal path compensation: Disconnect all input signals. Press
the main-menu button Signal Path; then press the side-menu button
OK Compensate Signal Paths.
NOTE
Failure to run the signal path compensation may result in the oscillo-
scope not meeting warranted performance levels.
g.
Wait: signal path compensation runs in one to three minutes. While it
progresses, a “clock” icon (shown at left) displays on-screen. When
compensation completes, the status message updates to
Pass or Fail
in the main menu (see step h).
h.
Confirm signal path compensation returns passed status: Verify the
word
Pass appears under Signal Path in the main menu. (See
Figure 4-2.)
Содержание 410
Страница 9: ...Contents x ...
Страница 13: ...Contents xx ...
Страница 19: ...Safety Summary xx ...
Страница 25: ...Introduction x x ...
Страница 26: ...Specifications ...
Страница 47: ...Operating Information ...
Страница 63: ...Operating Information Operating Information 2Ć16 ...
Страница 64: ...Theory of Operation ...
Страница 68: ...Circuit Description Theory of Operation 3Ć4 ...
Страница 69: ...Performance Verification ...
Страница 81: ...Brief Procedures Performance Verification 4Ć12 ...
Страница 119: ...Performance Tests Performance Verification 4Ć50 ...
Страница 120: ...Adjustment Procedures ...
Страница 147: ...Maintenance ...
Страница 151: ...Maintenance Information Maintenance 6Ć4 ...
Страница 228: ...Troubleshooting TDS 410 TDS 420 TDS 460 Service Manual 6Ć81 J2 Pin 1 Figure 6 38 A02 Display Connector J2 ...
Страница 237: ...Troubleshooting Maintenance 6Ć90 J7 Pin 1 J8 Pin 1 J6 Pin 1 W1 Pin 1 W2 Pin 1 Figure 6 46 A01 Backplane Module ...
Страница 240: ...Options ...
Страница 249: ...Electrical Parts List ...
Страница 251: ...Replaceable Electrical Parts List Electrical Parts List 8Ć2 ...
Страница 252: ...Diagrams ...
Страница 258: ...Diagrams 9Ć6 ...
Страница 259: ...Mechanical Parts List ...
Страница 263: ...Mechanical Parts List 10Ć4 ...
Страница 272: ...TDS 410 TDS 420 TDS 460 Service Manual 10Ć13 1 2 3 4 5 6 7 8 9 10 11 6 3 Figure 10 3 Cables and Routing ...
Страница 275: ...Mechanical Parts List 10Ć16 ...