Troubleshooting
Maintenance
6Ć70
Power off and remove the
cover using the
Front
Cover, Rear Cover,
Cabinet, Rear EMI Gasket,
and Cabinet Handle and
Feet removal procedure.
Check all the cables
coming out of the Low
Voltage Power Supply. and
the cabling between
modules. Be sure every
cable is attached securely.
Replace the
A03 CPU
module.
!
Note: The
replacement
module will not
have firmware
loaded. See 6-66
for details on
firmware
updates.
Perform the Console
troubleshooting procedure.
!
Does
the oscilloscope
power on and all the
diagnostics
pass?
Yes
No
Yes
No
No
Yes
Press on the principal power
switch on the back of the
oscilloscope.
Can you
hear the fan
whirling?
Press
the ON/STBY
button. Can you
hear the fan
now?
When
the oscilloscope
powers on, do the front
panel lights come on and
then go off a short
time later?
Yes
No
Perform the
Processor/Front Panel
troubleshooting procedure.
Does the
system power
on now?
Yes
Does
the display seem
to be working
at all?
No
Does
DS1 flash .
8
,
then display the
sequence of hex numbers
pausing to
flash .
b
?
Set both switches of S2 on the
A03 CPU module to the open
positions. Set all switches of S1
back to the closed position and
cycle power.
!
Is
the display
readable and
stable?
Yes
No
No
Yes
Yes
No
Perform the Monitor
troubleshooting
procedure.
!
There is something
wrong with the
ON/STBY button.
Perform the
Backplane
troubleshooting
procedure.
No
Yes
Power off the
oscilloscope and
remove its cabinet using
the
Front Cover, Rear
Cover, Cabinet, Rear
EMI Gasket, and
Cabinet Handle and
Feet removal procedure.
On
the A03 CPU module
set the eighth switch of
S1 to the open position.
Does DS1
first flash .
8
, then
display a sequence of
hex numbers from
1–e with no period
preceding
them?
Perform the Low
Voltage Power Supply
troubleshooting
procedure.
Perform the Module Isolation
troubleshooting procedure.
Does
the oscilloscope
respond correctly
when the front
panel buttons are
pushed?
Yes
No
Perform the
Processor/Front Panel
troubleshooting procedure.
Power
on again and
observe LED
DS1. Does it
only display
.
8
?
Yes
Replace the
A03 CPU
module.
!
Is
there 5.1 V
on P1 pin 3
(see figure
6-28)?
Yes
No
Perform the Low Voltage
Power Supply
troubleshooting procedure.
!
No
!
Note: Set all the switches of S1 back to the closed
position and cycle power before performing another procedure.
Done.
Do you
have a console
port?
No
Yes
Are both switches
of S2 in the open
position?
No
Yes
Figure 6-27: Primary Troubleshooting Procedure
Содержание 410
Страница 9: ...Contents x ...
Страница 13: ...Contents xx ...
Страница 19: ...Safety Summary xx ...
Страница 25: ...Introduction x x ...
Страница 26: ...Specifications ...
Страница 47: ...Operating Information ...
Страница 63: ...Operating Information Operating Information 2Ć16 ...
Страница 64: ...Theory of Operation ...
Страница 68: ...Circuit Description Theory of Operation 3Ć4 ...
Страница 69: ...Performance Verification ...
Страница 81: ...Brief Procedures Performance Verification 4Ć12 ...
Страница 119: ...Performance Tests Performance Verification 4Ć50 ...
Страница 120: ...Adjustment Procedures ...
Страница 147: ...Maintenance ...
Страница 151: ...Maintenance Information Maintenance 6Ć4 ...
Страница 228: ...Troubleshooting TDS 410 TDS 420 TDS 460 Service Manual 6Ć81 J2 Pin 1 Figure 6 38 A02 Display Connector J2 ...
Страница 237: ...Troubleshooting Maintenance 6Ć90 J7 Pin 1 J8 Pin 1 J6 Pin 1 W1 Pin 1 W2 Pin 1 Figure 6 46 A01 Backplane Module ...
Страница 240: ...Options ...
Страница 249: ...Electrical Parts List ...
Страница 251: ...Replaceable Electrical Parts List Electrical Parts List 8Ć2 ...
Страница 252: ...Diagrams ...
Страница 258: ...Diagrams 9Ć6 ...
Страница 259: ...Mechanical Parts List ...
Страница 263: ...Mechanical Parts List 10Ć4 ...
Страница 272: ...TDS 410 TDS 420 TDS 460 Service Manual 10Ć13 1 2 3 4 5 6 7 8 9 10 11 6 3 Figure 10 3 Cables and Routing ...
Страница 275: ...Mechanical Parts List 10Ć16 ...