Operating Instructions-Type 310A
2. Set the T IM E /D IV switch and the Variable T IM E /
D IV control so one cycle of the signal covers an exact
number of horizontal divisions. Do not change the Variable
T IM E /D IV control after obtaining the desired deflection.
This display can be used as a reference fo r frequency com
parison measurements.
3. To establish an arbitrary sweep rate so the repetition
rate of an unknown signal can be measured accurately at
any setting of the T IM E /D IV switch, the repetition rate of
the reference signal must be known. If it is not known, it
can be measured before the Variable T IM E /D IV switch is
set in step 2.
in seven horizontal divisions can be determined by using the
repetition rate formula (step 6):
Repetition
Rate
= 50ps X 1.37 X 7
480 jus
This answer can be converted to frequency by taking the
reciprocal o f the repetition rate (see application on Deter
mining Frequency).
Time-Duration Measurements
To measure time between tw o points on a waveform, use
the follow ing procedure:
4.
Divide the repetition rate o f the reference signal (sec
onds) by the product o f the horizontal deflection estab
lished in step 2 (divisions) and the setting of the T IM E /D IV
switch. This is the horizontal conversion factor.
1. Connect the signal to the vertical INPUT connector.
2. Set the V O LTS/D IV switch to display about fo ur
divisions o f the waveform.
reference signal repetition rate (seconds)
horizontal
T IM E /D IV
deflection X
switch
(divisions)
setting
5. To measure the repetition rate o f an unknown signal,
disconnect the reference signal and connect the unknown
signal to the vertical INPUT connector. Set the T IM E /D IV
switch to a setting that provides sufficient horizontal de
flection to make an accurate measurement. Do not readjust
the Variable T IM E /D IV control.
6. Measure the horizontal deflection in divisions and
calculate the repetition rate of the unknown signal using
the following formula:
Horizontal
Conversion =
Factor
Repetition
Rate
T IM E /D IV
horizontal
horizontal
switch
X conversion X
deflection
setting
factor
(divisions)
NOTE
If the horizontal magnifier is used, be sure to use the
magnified sweep rate in place of the T IM E /D IV
switch setting.
3. Set the triggering control to obtain a stable display.
4. Set the T IM E /D IV switch to the fastest sweep rate
that displays less than eight divisions between the time
measurement points (see Fig. 2-11). (See the topic entitled
Horizontal Sweep Rate in this section concerning non
linearity o f first and last divisions o f display.)
Fig. 2 -1 1 . Measuring the tim e du ratio n between points on a wave-
form .
Example.
Assume a reference signal frequency of 455
hertz (repetition rate 2.19 milliseconds), a T IM E /D IV
switch setting o f .2 ms and the Variable T IM E /D IV control
is adjusted to provide a horizontal deflection o f eight d ivi
sions. Substituting these values in the horizontal conversion
factor form ula (step 4):
Horizontal
Conversion =
Factor
2.19 milliseconds
. 2 X 8
1.37
Then, w ith a T IM E /D IV switch setting of 50 [xs, the repe
titio n rate of an unknown signal which completes one cycle
5. Adjust the V E R TIC A L POSITION control to move
the points between which the time measurement is made to
the center horizontal line.
6. Adjust the HO RIZO NTAL POSITION control to cen
ter the time-measurement points w ithin the center eight
divisions o f the graticule.
7. Measure the horizontal distance between the time
measurement points. Be sure the V A R IA B LE T IM E /D IV
control is set to C ALIBRATED.
8. M ultiply the distance measured in step 7 by the set
ting of the T IM E /D IV switch. If sweep magnification is
used, divide this answer by 5.
2-17
Содержание 310A
Страница 3: ...Type 310A ...
Страница 8: ...NOTES ...
Страница 10: ...Operating Instructions Type 310A 2 2 Fifl 2 2 Front and rear panel controls and connectors ...
Страница 16: ...rsj 60 T Y P E 3 1 0 A T E S T S E T U P C H A R T Fig 2 3 Operating Instructions Type 310A ...
Страница 28: ...NOTES ...
Страница 41: ...Circuit Description Type 310A SW 600 POWER I 3 13 Fig 3 8 LV Power Supply detailed block diagram ...
Страница 44: ...NOTES ...
Страница 88: ...SECTION 7 M ECHANICAL PARTS LIST Type 310A ...
Страница 98: ...Mechanical Parts List Type 310A POWER CHASSIS 7 1 0 ...
Страница 102: ...Mechanical Parts List Type 310A VERTICAL AMPLIFIER CHASSIS 7 1 4 ...
Страница 106: ...Mechanical Parts List Type 310A REAR 7 1 8 ...
Страница 112: ...S Q U A R E W A V E C A LIB R A TO R CAL OUT V50I V52Q T Y P E 3 I O A O S C I L L O S C O P E ...
Страница 113: ...V 8 59 B L O C K D I A G R A M ...
Страница 116: ...0 ...
Страница 118: ...I 1 SW 75A SW 75B r 5W 75C f T il TO CATH PIN J VI60A SWEEP GEN C F SWEEP TYPE 3 0A OSCILLOSCOPE sw ...
Страница 119: ... 1 TIM E DIV 1 SW 75C SWI75D I t ...
Страница 122: ...I 1 1 1 1 1 V0LTS 3W 30SA 3W 30SB 3W 305C 3W 305D 3W303C 3 W 3 0 5 F T Y P E 3 0A O SC ILLO SCO PE ...
Страница 124: ...V40 V40a INPUT AM PLIFIER S V430t V430L DRIVER C DECOUPLING N E TW O R K S T Y P E 3 0 A OSCILLOSCOPE ...
Страница 129: ... CRT CIRCUIT C IR C U IT NUMBfcRS 7 0 0 THRU 7 6 0 ...
Страница 131: ...CALIB R A TO R CATHODE FO LLO W E R IOOV CALIBRATOR c C IR C U IT NUMBfep S SOO THRU 5 4 9 ...