Table 4 {cont)
Characteristics
Performance Requirements
Supplemental Inform ation
VE R TIC A L DEFLECTION SYSTEM (cont)
Input Gate Current
—15°C to +30° C
0.5 nA or less.
0.1-division trace shift when moving
Input Coupling switch from GND to
AC at 5 mV per division.
+30° C to +55° C
4.0 nA or less.3
0.8-division trace shift when moving
input Coupling switch from GND to
AC at 5 mV per division.
Attenuator Isolation (CH 1 to CH 2)
A t least 100 to 1.
With one vertical input set at 0.5 V per
division, apply 4-V p-p 25-MHz signal;
set the other vertical input to 10 mV
per division. Check for less than
4 divisions of signal.
POSITION Control Range
A t least +12 and —12 divisions from
graticule center.
Perform ance requirement not checked in manuat.
2335 Operators
69
Содержание 2335
Страница 1: ...Tektronix Inc P O Box 500 Beaverton Oregon 97077 070 4115 00 ...
Страница 11: ......
Страница 22: ...4115 Figure 5 Horizontal controls and indicator ...
Страница 36: ...AMPL OUTPUT 4 115 26 Figure 9 initial setup for instrument familiarization procedure 2335 Operators 23 ...
Страница 68: ...INTENSIFIED ZONE A A INTENSIFIED TRACE 4 1 1 5 2 0 Figure 21 Delayed sweep magnification ...
Страница 103: ......
Страница 104: ...NOTES ...