4. Establish an arbitrary deflection factor, using the
follow ing formual:
Substituting the given values in the horizontal conver
sion factor formula:
A rb itra ry
horizontal
A SEC/DIV
Deflection = conversion x
switch
Factor
factor
setting
Horizontal
Conversion
Factor
_____ 2.19 ms_____
8 d iv x 0.2 ms/div
1.37
5. Measure the horizontal distance of the unknown
signal in divisions and calculate its time duration using the
follow ing form ula:
Continuing, fo r the unknown signal the A SEC/DIV
switch setting is 50 jus, and one complete cycle spans 7
horizontal divisions. The arbitrary deflection factor is then
determined by substituting values in the formula:
Time
Duration
arbitrary
deflection
factor
horizontal
x
distance
(divisions)
Arbitrary
Deflection =
1.37 x 50 jus/div = 68.5 jus/div
Factor
6. Frequency of the unknown signal can then be deter
mined by calculating the reciprocal of its time duration.
The time duration o f the unknown signal can then be
computed by substituting values in the formula:
T
' m o
.
= 68.5 jus/div x 7 div = 480 jus/div
Duration
EXAMPLE: The reference signal time duration is
2.19 ms, the A SEC/DIV switch setting is 0.2 ms, and
the TIM E (PULL) VAR control is adjusted to provide
a horizontal distance o f exactly 8 divisions.
The frequency o f the unknown signal is then calculated:
Frequency
=
— =
2.083 kHz
480 jus
52
2335 Operators
Содержание 2335
Страница 1: ...Tektronix Inc P O Box 500 Beaverton Oregon 97077 070 4115 00 ...
Страница 11: ......
Страница 22: ...4115 Figure 5 Horizontal controls and indicator ...
Страница 36: ...AMPL OUTPUT 4 115 26 Figure 9 initial setup for instrument familiarization procedure 2335 Operators 23 ...
Страница 68: ...INTENSIFIED ZONE A A INTENSIFIED TRACE 4 1 1 5 2 0 Figure 21 Delayed sweep magnification ...
Страница 103: ......
Страница 104: ...NOTES ...