13.
Observe that the TRIG 'D-READY light illuminates,
indicating that the A trigger circuit is armed (READY)
fo r a single-sweep display. No display should be present
on the crt screen.
14,
Reconnect the calibration signal to the CH 1 input
connector. A single sweep o f the applied signal should
appear on the screen. When the TR IG 'D -R EAD Y light is
out, another single sweep cannot be displayed until the
SGL SWP button is pressed in again to reset the A Trigger
circuit.
15.
Select the A TRIGGER AUTO mode and set the
A TRIGGER SOURCE switch to EXT.
16.
Remove the calibration signal from the CH 2 input
connector and connect it to the A EXT inp ut connector.
17.
Set the Channel 1 VO LTS/DIV switch to 0.5 (IX )
and adjust the output o f the calibration generator to pro
vide a 4-division display. Adjust the A TRIGGER LEVEL
control fo r a stable display.
18. Set the A TRIGGER SOURCE switch to E X T-M O .
19.
Observe th a t adjustment o f the A TRIGGER LEVEL
control provides a triggered display over a narrower
range than in preceding step 17, indicating trigger-signal
attenuation.
Using the X-Y Mode
1. Remove the calibration signal from the A EXT input
connector and reconnect it to the CH 2 input connector.
Set the A TRIGGER SOURCE switch to VERT MODE and
adjust the A TRIGGER LEVEL control fo r a stable display.
2. Set both the CH 1 and CH 2 VO LTS/DIV switches
to 1 (IX ) and adjust the generator output to provide a
5-division display.
3.
Select X-Y VER T MODE by simultaneously pressing
in the CH 1 and CH 2 push buttons.
4.
Increase the INTEN control setting until tw o dots
are displayed diagonally. This display can then be posi
tioned horizontally w ith the Horizontal POSITION control
and vertically w ith the Channel 2 POSITION control.
Note that the dots are separated by 5 horizontal divisions
and 5 vertical divisions.
30
2335 Operators
Содержание 2335
Страница 1: ...Tektronix Inc P O Box 500 Beaverton Oregon 97077 070 4115 00 ...
Страница 11: ......
Страница 22: ...4115 Figure 5 Horizontal controls and indicator ...
Страница 36: ...AMPL OUTPUT 4 115 26 Figure 9 initial setup for instrument familiarization procedure 2335 Operators 23 ...
Страница 68: ...INTENSIFIED ZONE A A INTENSIFIED TRACE 4 1 1 5 2 0 Figure 21 Delayed sweep magnification ...
Страница 103: ......