Table 4 (cont)
Characteristics
Performance Requirements
Supplemental Information
VERTICAL DEFLECTION SYSTEM (cont)
Step Response (cont)
Rise Time (5 mV per division to
5 V per division)
— 15° C to +40° C
3.5 ns or less.
Rise time is calculated from the
formula:
"f.
0.35
Rise Time -
.. . . . . . .
BW (in MHz)
+40° C to +55° C
4.15 ns or less.3
Aberrations
Positive-Going Step
(Excluding ADD Mode)
5 mV per division to 0.2 V
per division
+3%,
—3%,
3% p-p or less.
Negative-Going Step
Add 2% to all positive-going step
specifications; checked at 5 mV
per division.
Performance requirement not checked in manual.
2335 Operators
67
Содержание 2335
Страница 1: ...Tektronix Inc P O Box 500 Beaverton Oregon 97077 070 4115 00 ...
Страница 11: ......
Страница 22: ...4115 Figure 5 Horizontal controls and indicator ...
Страница 36: ...AMPL OUTPUT 4 115 26 Figure 9 initial setup for instrument familiarization procedure 2335 Operators 23 ...
Страница 68: ...INTENSIFIED ZONE A A INTENSIFIED TRACE 4 1 1 5 2 0 Figure 21 Delayed sweep magnification ...
Страница 103: ......
Страница 104: ...NOTES ...