2. Apply the signal to either vertical-channel input
EXAMPLE: The distance between the time-measurement
connector and set the VE R TIC A L MODE switch to display
points is 8 divisions (see Figure 16), and the A SEC/DIV
the channel used.
switch is set to 2 ms. A magnification factor of 1 is used.
3.
Adjust the A TRIGGER LEVEL control to obtain
a stable display.
4.
Set the A SEC/DIV switch to display one complete
period o f the waveform. Ensure that the TIME (PULL)
VAR control is in the calibrated detent.
5.
Position the display to place the time-measurement
points on the center horizontal graticule line (see Fig
ure 16).
6.
Measure the horizontal distance between the time-
measurement points.
7. Calculate time duration, using the following formula:
horizontal
A SEC/DIV
distance
x
switch
Time
_
(divisions)_______ setting
Duration
magnification factor
Substituting the given values:
Time Duration = 8.3 div x 2 ms/div = 16.6 ms
2335 Operators
43