Time Difference Between Two Time-Related Pulses
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Using probes or cables having equal time delays,
apply the reference signal to the Channel 1 input and apply
the comparison signal to the Channel 2 input.
3.
Set both VO LTS/DIV switches to produce a display
of either 4 or 5 divisions in amplitude.
4.
Select either A L T or CHOP V E R T IC A L MODE,
depending on the input frequencies (or select AUTO
VE R TIC A L MODE if automatic selection is desired).
5.
Set the A SEC/DIV switch to display the measure
ment points o f interest w ithin the graticule area.
6.
Select the A INTEN HORIZ MODE and set the
B SEC/DIV switch to the fastest sweep speed that provides
a usable (visible) intensified zone.
7. Adjust the B D ELAY TIME POSITION control to
move the intensified zone to the first pulse (see Figure 27A,
Point A).
8. Select the B HORIZ MODE and adjust the B D ELAY
TIME POSITION control to move the rising portion o f the
first pulse (reference signal) to a convenient vertical refer
ence line (see Figure 27B).
9.
Record the B D ELAY TIME POSITION control dial
setting.
10.
Adjust the B D ELAY TIME POSITION control
clockwise until the rising portion o f the second pulse
(comparison signal) is positioned to the same vertical refer
ence line (see Figure 27B).
NOTE
I f several pulses are displayed, return to the A INTEN
position to locate the correct pulse. Do not change
the setting o f the horizontal POSITION control.
11.
Record the B D ELAY TIME POSITION control
dial setting.
@
2335 Operators
81
Содержание 2335
Страница 1: ...Tektronix Inc P O Box 500 Beaverton Oregon 97077 070 4115 00 ...
Страница 11: ......
Страница 22: ...4115 Figure 5 Horizontal controls and indicator ...
Страница 36: ...AMPL OUTPUT 4 115 26 Figure 9 initial setup for instrument familiarization procedure 2335 Operators 23 ...
Страница 68: ...INTENSIFIED ZONE A A INTENSIFIED TRACE 4 1 1 5 2 0 Figure 21 Delayed sweep magnification ...
Страница 103: ......
Страница 104: ...NOTES ...