Calibration— Type ! 130
U
*
(B) Attenuator arrangement.
I
Fig. 6-20. Test equipment setup to check IF ATTEN dB accuracy, dynamic range, IF GAIN control range, RECORDER out signal level and VIDEO
FILTER operation. (Steps 9 through 13.)
Type 1L30
Plug-in Oscilloscope
POS
IF CENTER FREQ
FINE
DISPERSION RANGE
DISPERSION-COUPLED
RESOLUTION
VIDEO FILTER
G A IN
VERTICAL DISPLAY
IF ATTEN
FINE RF CENTER FREQ
INT 1 M H z REF FREQ
Position a free running
trace, to the bottom
line of the graticule
Midrange (000)
Centered
kH z/C M
500 kHz
OFF
Midrange
LIN
Switches off
Centered
OFF
©
Time/Cm
10 ms
Triggering
Adjusted for a free
running sweep
Horizontal Position
Centered trace
9. Check Accuracy of IF Attenuator dB Selectors
a. Equipment setup is shown in Fig. 6-20.
b. Apply a 200 M Hz signal, that is 10 dB below 1 mW,
from the signal generator through a 2 X Attenuator (6 dB),
a Tens and Units Step Attenuator and a 1 0 X Attenuator
(20 dB) to the Type 1L30 RF INPUT connector.
c. Set the Tens Step Attenuator for -20 dB and -the Units
Step Attenuator for 12 dB attenuation.
6 -1 7