Calibration— Type 1L30
Fig. 6-2. Recommended test equipment to check dial calibration, analyzer sensitivity and response flatness.
18. Adapter, BNC female to N male; Tektronix Part No.
103-0045-00.
19. Miniature phone plug with 600 Q, load. (Test fixture
to check TO RECORDER signal amplitude.) Consist of a 600 f2
5% V
2
watt resistor, soldered across a miniature phone plug.
20. Two (2) adapters. Sealectro to GR: Type P6040 probe
cable, Part No. 010-0133-00; or Sealectro to BNC, Type
P6041 probe; Part No. 010-0164-00.
21. Two (2) BNC, coaxial cables, 50 Q. Tektronix Part No.
012-0057-01.
22. Patch cord with BNC to banana plug tips; Tektronix
Part No. 012-0091-00.
23. Flexible Cable Plug-In Extension; Tektronix Part No.
012-0038-00.
24. Adjusting tools:
a. Screwdriver, 3/ 32 blade, 3 inch shaft
003-0192-00
b. Tuning tool
Handle
003-0307-00
Insert for
% 4
(D) hex cores
003-0310-00
c. Low capacitance screwdriver, V4 inch
003-0209-00
by 8 inch fiber rod with screwdriver
shaped ends
Group Two
RF Signal Generators with calibrated frequency and out
put power: Frequency range 925 M Hz to 10,500 MHz, ac
curacy ± 1 % ; output power — 100 dBm to — 30 dBm; output
impedance 50 fi. Suggested equipment:
Hewlett-Packard 8614A UHF signal generator, 800 M Hz
to 2400 M Hz.
Hewlett-Packard 8616A UHF signal generator, 1800 M H z
to 4500 M Hz.
Polarad Type 1107 Microwave signal generator 3.8 G Hz
to 8.2 GHz.
Polarad Tyype 1108 Microwave signal generator 6.95 G H z
to 11,0 GHz.
1
6
©
Содержание 1L30
Страница 48: ...si L i y n Maintenance Type 1L30 Fig 4 15 Honeycomb assembly drcui a n d component layout 4 n ...
Страница 59: ...Fig 6 1 A Test equipment required for calibration ...
Страница 60: ... Calibration Type 1130 ...
Страница 120: ...T Y P E I L 3 0 S P E C T R U M A N A L Y Z E R ib i IF SYSTEM BLOCK DIAGRAM 9 1 9 2 ...
Страница 127: ......