d. Adjust the G A IN control for a signal amplitude of
4 cm on the plug-in oscilloscope screen.
e. Check the accuracy of the IF ATTEN dB selectors as
follows:
1. Switch the Type 1130 1 dB attenuator switch to
O N and switch out 1 dB of attenuation through the units
Step Attenuator.
2. Check the display amplitude.
Must equal 4 cm,
± 0 . 5 mm (.1 dB/dB).
3. Switch the Type 1L30 IF ATTEN switch to OFF posi
tion, then check the remaining IF ATTEN switch steps
as directed in Table 5-3a.
TABLE 5 - 3 a
Spectrum A nalyzer
IF ATTEN Switch
On
Step
Attenuators
Signal A m plitude
Limits (.1 d B /d B )
Units
Tens
1 dB
11
20
3.95 cm to 4.05 cm
2 dB
10
20
3.9 cm to 4.1 cm
4 dB
8
20
3.8 cm to 4.2 cm
. 8 dB
4 .
20
3.6 cm to 4.4 cm
16 dB
6
10
3.2 cm to 4.8 cm
20 dB
2
10
2.95 cm to 5.05 cm
The 1 and 2 dB measurements are very difficult, because
of signal stability and the noise level. For these small sig
nal levels, the square law mode may be used to expand the
screen changes for the same level change by the square
power as listed in Table 5-3b.
|
TABLE 5 -3 b
dB
1
2
4
8
16
20
Signal
3.95
3.9
3.6
3.2
2.2
1.7
Amplitude
to
to
to
to
to
to
limits
4.05
4.1
4.4
4.8
5.8
6.3
An alternate method which is not as accurate but is suf
ficient for most applications is as follows:
1. Apply a 200 MHz signal (at 60 dB below 1 mW, as
shown on the Attenuator dial) from the Signal Generator to
the R.r INPUT connector. Adjust the Spectrum Analyzer G A IN
control for a signal amplitude of 5 cm.
5
-
6
-
2. Switch the 1 dB IF ATTEN switch on and adjust the
Signal Generator output attenuator control to return the
signal amplitude to 5 cm.
3. Check the new reading of the attenuator dial. Should
read — 59 dBm ± 0 .1 dBm.
4. Turn the 1 dB IF ATTEN switch to OFF. Check the re
mainder of the IF ATTEN selector steps as directed in
Table 5-3c.
TABLE 5 -3 c
Spectrum A n a ly z e r
IF ATTEN switch on
RF G e n e ra to r Attenuatjor
Control Setting
2 dB
— 58 dBm ± . 2 dBm
4dB
— 56 dBm ± . 4 dBm
8 dB
— 52 dBm ± . 8 dBm
16 dB
— 44 dBm ± 1 . 6 dBm
20 dB
— 40 dBm ± 2 . 0 dBm
9. Check Attenuation Range of G A IN Control
a. Requirement— The attenuation range of the G A IN con
trol should equal or exceed 50 dB.
b. W ith the 200 M Hz IF feedthrough signal applied as
directed in step 7 and the G A IN control turned fully counter
clockwise, adjust the Signal Generator Variable Attenuator
control for a signal amplitude of 6 cm. Note the Attenuator
reading in dBm.
c. Increase the Signal Generator output attenuation by
50 dB. Turn the G A IN control fully clockwise.
d. Check— Signal amplitude must equal or exceed 6 cm.
(Range > 5 0 dB.) If the range does not meet this requirement,
recheck the adjustment of the Narrow Band IF, as described
in the Calibration section.
10. Check Signal Amplitude at TO RECORDER
Connector
a. Requirement— Signal amplitude at the TO RECORDER
output connector with a full screen display (6 cm) should
measure -between 12 mV and 20 mV when the output is
terminated into a 600 ohm load.
b. Set the front panel controls as follows:
©
Содержание 1L30
Страница 48: ...si L i y n Maintenance Type 1L30 Fig 4 15 Honeycomb assembly drcui a n d component layout 4 n ...
Страница 59: ...Fig 6 1 A Test equipment required for calibration ...
Страница 60: ... Calibration Type 1130 ...
Страница 120: ...T Y P E I L 3 0 S P E C T R U M A N A L Y Z E R ib i IF SYSTEM BLOCK DIAGRAM 9 1 9 2 ...
Страница 127: ......