Performance Verification
6–16
1710J-Series Waveform Monitors
Spectrum analyzer
TG
OUTPUT
RF
INPUT
75
W
Precision
terminator
High frequency
RF bridge
RF Out
RF In
Figure 6–5: Connecting the RF bridge to the spectrum analyzer and tracking
generator
3. Select Demod/TG on the spectrum analyzer. Turn on the tracking generator
and set the tracking generator fixed level to 0.00 dBm.
4. Set the spectrum analyzer Span/Div to 1 MHz and the Resolution Bandwidth
to 300 kHz.
5. Set the spectrum analyzer Vertical Scale to 10 dB.
6. Set the spectrum analyzer Reference Level to the first major division down
from the top on the analyzer display.
7. Remove one of the cables from the RF bridge.
8. Set the spectrum analyzer Frequency to 5 MHz and turn the Marker on. Set
the Marker to 6 MHz.
9. Reconnect the cable to the RF bridge.
10. Observe the Reference Level Readout.
11. Adjust the spectrum analyzer External Attenuation Amplitude (on the
spectrum analyzer Input menu) by the amount noted in step 10. The
Reference Level Readout should now be 0.00 dBm.
12. Connect the precision high-frequency terminator to the Device Under Test
connector on the RF bridge.
13. Check that the frequency response from 0 MHz to 6 MHz is
≥
40 dBm.
14. Return the spectrum analyzer frequency marker to 6 MHz if it was moved.
15. Remove the precision high-frequency terminator from the RF bridge.
Содержание 1710J Series
Страница 4: ......
Страница 14: ...Table of Contents x 1710J Series Waveform Monitors ...
Страница 18: ...Service Safety Summary xiv 1710J Series Waveform Monitors ...
Страница 22: ...Preface xviii 1710J Series Waveform Monitors ...
Страница 23: ...Getting Started ...
Страница 24: ......
Страница 45: ...Operating Basics ...
Страница 46: ......
Страница 55: ...Reference ...
Страница 56: ......
Страница 64: ...Reference 3 8 1710J Series Waveform Monitors ...
Страница 65: ...Specifications ...
Страница 66: ......
Страница 74: ...Specifications 4 8 1710J Series Waveform Monitors ...
Страница 76: ......
Страница 77: ...Theory of Operation ...
Страница 78: ......
Страница 100: ...Theory of Operation 5 22 1710J Series Waveform Monitors ...
Страница 101: ...Performance Verification ...
Страница 102: ......
Страница 120: ...Performance Verification 6 18 1710J Series Waveform Monitors ...
Страница 121: ...Adjustment Procedure ...
Страница 122: ......
Страница 138: ...Adjustment Procedure 7 16 1710J Series Waveform Monitors ...
Страница 139: ...Maintenance ...
Страница 140: ......
Страница 168: ...Maintenance 8 28 1710J Series Waveform Monitors Figure 8 11 Repackaging a 1710J Series Waveform Monitor ...
Страница 169: ...Replaceable Electrical Parts ...
Страница 170: ......
Страница 204: ...Replaceable Electrical Parts 9 34 1710J Series Waveform Monitors ...
Страница 205: ...Diagrams ...
Страница 206: ......
Страница 226: ......
Страница 227: ...Replaceable Mechanical Parts ...
Страница 228: ......
Страница 239: ...Replaceable Mechanical Parts 1710J Series Waveform Monitors 11 11 1 2 3 4 5 6 Figure 11 4 Accessories ...
Страница 240: ...Replaceable Mechanical Parts 11 12 1710J Series Waveform Monitors ...
Страница 244: ...Glossary Glossary 4 1710J Series Waveform Monitors ...
Страница 251: ......
Страница 252: ......