Performance Verification
6–12
1710J-Series Waveform Monitors
Using 50 kHz as the reference the requirements are
H
X1 gain 250 kHz to 6 MHz within 2%
H
X5 gain 250 kHz to 6 MHz within 5%
1. Connect a 50
Ω
precision cable and 50 to 75
Ω
minimum loss pad from the
leveled sine wave generator to the CH A input. Terminate the remaining side
of the loop-through input with a 75
Ω
end-line terminator. See Figure 6–4.
1710J-Series Waveform Monitor
75
W
Terminator
Leveled sine wave Generator
to Test signal
Black Burst signal
Figure 6–4: Initial equipment setup to check Flat Response
2. Select EXT REF.
3. Set the leveled sine wave generator frequency range to 50 kHz and set the
output amplitude for a 100 IRE (0.7 V) display amplitude.
4. Check the flat response using the 50 kHz response as a reference. While
changing the frequency of the sine wave generator, check that the response is
within
±
2% from 250 kHz to 6 MHz.
5. Change the VERTICAL GAIN to X5.
6. Set the leveled sine wave generator to 50 kHz and set the output amplitude
for a 100 IRE (0.7 V) display amplitude.
7. Check flat response using the 50 kHz response as a reference. Check that the
amplitude is 100 IRE (0.7 V)
±
5% from 250 kHz to 6 MHz.
8. Move the 75
W
coaxial cable that comes from the leveled sine wave
generator output to CH B. Terminate the remaining side of the loop-through
input with a 75
W
end-line terminator.
9. Select CH B input.
10. Check flat response using the 50 kHz response as a reference. Check that the
amplitude is 100 IRE (0.7 V)
±
5% from 250 kHz to 6 MHz.
Check Flat Response
Содержание 1710J Series
Страница 4: ......
Страница 14: ...Table of Contents x 1710J Series Waveform Monitors ...
Страница 18: ...Service Safety Summary xiv 1710J Series Waveform Monitors ...
Страница 22: ...Preface xviii 1710J Series Waveform Monitors ...
Страница 23: ...Getting Started ...
Страница 24: ......
Страница 45: ...Operating Basics ...
Страница 46: ......
Страница 55: ...Reference ...
Страница 56: ......
Страница 64: ...Reference 3 8 1710J Series Waveform Monitors ...
Страница 65: ...Specifications ...
Страница 66: ......
Страница 74: ...Specifications 4 8 1710J Series Waveform Monitors ...
Страница 76: ......
Страница 77: ...Theory of Operation ...
Страница 78: ......
Страница 100: ...Theory of Operation 5 22 1710J Series Waveform Monitors ...
Страница 101: ...Performance Verification ...
Страница 102: ......
Страница 120: ...Performance Verification 6 18 1710J Series Waveform Monitors ...
Страница 121: ...Adjustment Procedure ...
Страница 122: ......
Страница 138: ...Adjustment Procedure 7 16 1710J Series Waveform Monitors ...
Страница 139: ...Maintenance ...
Страница 140: ......
Страница 168: ...Maintenance 8 28 1710J Series Waveform Monitors Figure 8 11 Repackaging a 1710J Series Waveform Monitor ...
Страница 169: ...Replaceable Electrical Parts ...
Страница 170: ......
Страница 204: ...Replaceable Electrical Parts 9 34 1710J Series Waveform Monitors ...
Страница 205: ...Diagrams ...
Страница 206: ......
Страница 226: ......
Страница 227: ...Replaceable Mechanical Parts ...
Страница 228: ......
Страница 239: ...Replaceable Mechanical Parts 1710J Series Waveform Monitors 11 11 1 2 3 4 5 6 Figure 11 4 Accessories ...
Страница 240: ...Replaceable Mechanical Parts 11 12 1710J Series Waveform Monitors ...
Страница 244: ...Glossary Glossary 4 1710J Series Waveform Monitors ...
Страница 251: ......
Страница 252: ......