Maintenance
1710J-Series Waveform Monitors
8–27
5. Measure the voltage at the anode end of CR3. It should be 50–150 V more
negative than the reading from the anode of CR4.
Repackaging
If the instrument is to be shipped to a Tektronix Service Center for service or
repair, attach a tag to the instrument showing:
1. Owner (with complete address) and the name of the person at your firm that
can be contacted.
2. Instrument serial number and a description of the service required.
Repackage the instrument in the original manner to provide adequate protection
(see Figure 8–11). If the original packaging is not available or is unfit for use,
repackage the instrument as follows:
1. Obtain a corrugated cardboard carton whose inside dimensions are at least
six inches greater than the dimensions of the instrument to allow room for
cushioning. The shipping carton should have a test strength of at least 275
pounds.
2. Surround the instrument with polyethylene sheeting to protect the finish.
3. Cushion the instrument on all sides by tightly packing dunnage or urethane
foam between the carton and the instrument. Allow three inches on all sides
for cushioning.
4. Seal the carton with shipping tape or an industrial stapler.
Identification Tag
Repackaging for Shipment
Содержание 1710J Series
Страница 4: ......
Страница 14: ...Table of Contents x 1710J Series Waveform Monitors ...
Страница 18: ...Service Safety Summary xiv 1710J Series Waveform Monitors ...
Страница 22: ...Preface xviii 1710J Series Waveform Monitors ...
Страница 23: ...Getting Started ...
Страница 24: ......
Страница 45: ...Operating Basics ...
Страница 46: ......
Страница 55: ...Reference ...
Страница 56: ......
Страница 64: ...Reference 3 8 1710J Series Waveform Monitors ...
Страница 65: ...Specifications ...
Страница 66: ......
Страница 74: ...Specifications 4 8 1710J Series Waveform Monitors ...
Страница 76: ......
Страница 77: ...Theory of Operation ...
Страница 78: ......
Страница 100: ...Theory of Operation 5 22 1710J Series Waveform Monitors ...
Страница 101: ...Performance Verification ...
Страница 102: ......
Страница 120: ...Performance Verification 6 18 1710J Series Waveform Monitors ...
Страница 121: ...Adjustment Procedure ...
Страница 122: ......
Страница 138: ...Adjustment Procedure 7 16 1710J Series Waveform Monitors ...
Страница 139: ...Maintenance ...
Страница 140: ......
Страница 168: ...Maintenance 8 28 1710J Series Waveform Monitors Figure 8 11 Repackaging a 1710J Series Waveform Monitor ...
Страница 169: ...Replaceable Electrical Parts ...
Страница 170: ......
Страница 204: ...Replaceable Electrical Parts 9 34 1710J Series Waveform Monitors ...
Страница 205: ...Diagrams ...
Страница 206: ......
Страница 226: ......
Страница 227: ...Replaceable Mechanical Parts ...
Страница 228: ......
Страница 239: ...Replaceable Mechanical Parts 1710J Series Waveform Monitors 11 11 1 2 3 4 5 6 Figure 11 4 Accessories ...
Страница 240: ...Replaceable Mechanical Parts 11 12 1710J Series Waveform Monitors ...
Страница 244: ...Glossary Glossary 4 1710J Series Waveform Monitors ...
Страница 251: ......
Страница 252: ......