Technology Solutions
Instruction Manual
Tek-Clamp 1200A-100H
www.tek-trol.com
16
4.4.3
Menu Window Details
Menu Window
No.
Function
Flow
rate/flow
totalizer
display
M00
Display instant flow rate/net totalizer. Adjust the units in M30-
M32
M01
Display instant flow rate/instant flow velocity. Adjust the units
in M30-M32
M02
Display instant flow rate/positive totalizer. Adjust the units in
M30-M32
M03
Display instant flow rate/negative totalizer. Adjust the units in
M30-M32
M04
Display instant flow rate/date time
M05
Display heat flow rate/total heat quantity. Adjust the units in
M84, M88
M06
Display temperature input T1, T2
M07
Display present battery voltage
M07
Display analogue input AI3, AI4
M08
Display system error code
M09
Display today net totalizer
Initial setup
M10
Input outside perimeter of pipe
M11
Input pipe outer diameter, data range:0-18000mm
M12
Input pipe wall thickness
M13
Input pipe inner diameter
M14
Choose the kinds of pipe materials
M15
Input sound velocity of pipe material
M16
Choose kinds of liner
M17
Input the sound velocity of liner
M18
Input the thickness of liner
M19
Input inner pipe wall absolute degree of roughness
M20
Choose kinds of fluids
M21
Input fluid velocity
M22
Input fluid viscosity
M23
Choose the types of transducers, including more than 20 types
to use
M24
Choose transducer installation method
M25
Display transducer installation space
M26
Parameter solidifying and setup
M27
Store and read installation parameters on installation point
M28
When signal set is turning poor, the transmitter keeps the last
data. Choosing ‘yes’ means when the signal is turning poor, the
flow meter will display last correct measured data