Instruction Manual
Tek-Clamp 1200A-100H
www.tek-trol.com
9
3.1.2
Clamp on mounting bracket transducer
•
Install Tek-Clamp 1200A - 100H on the pipe, to finish the flow measurement. The
need for cutting pipe gets eliminated.
•
Several types of bracket Tek-Clamp 1200A - 100H are used to measure pipe size
from 1/2’’ to 28’’.
•
Several types of bracket Tek-Clamp 1200A - 100H are used to measure
temperature range from -22°F to 320°F (-30°C to 160°C)
3.2
Requirements for Measuring Flow meter Accuracy
•
Medium must be conductive.
•
The pipe should be completely filled with the medium.
•
Medium conductivity should be homogeneous to avoid severe disruption. If chemical
substances need be injected into the pipe, the injection should be operated at upstream
side of flow meter.
•
Magnetic flow meter system should be connected to the ground.
•
There must be a straight pipe run no less than 5DN at flow meter upstream and 2DN at
flow meter downstream. (DN is the internal diameter of the pipeline).
•
Do not install the flow meter near large electrical machine or transformer in order to
avoid the electromagnetic interference.
NOTE
Do not expose the electronic unit box to direct sunlight; in case of outdoor
installation, use a suitable protection shield.
Fig 6: Tek-Clamp 1200A-100H Handheld Ultrasonic Clamp-on mounting bracket transducer