24
For pre-cleaning purposes, disconnect pressure remove the
dispensing fluid, then connect an empty but clean syringe to the
valve. Use the pressurised air hook up to the syringe to push out any
residue fluid left in the valve out.
The cleaning of the TS9800 Jet Valve can be carried out via:
•
Purge fluid path with appropriate cleaning solutions
To remove the remaining dispensed fluid, purge the system with
an appropriate cleaning solution (refer to the material safety
data sheet of the dispensed media and suitable cleaning
solution). For this method, connect a syringe filled with an
appropriate cleaning solution to the valve, then connect the
syringe to the supply pressure. Touch and hold the ‘Purge’ icon
on the controller to purge the system. Run the purge until
the system is clean. Ideally the only material left during the
purging should be the cleaning solution, that would be a good
indicator that the fluid path is clean.
•
Thorough Cleaning
1.
Perform the above step to purge the system with an appropriate
cleaning solution. Turn the system Off and disconnect all the
connections to the valve.
2.
Remove the nozzle adjustment nut (8) by using the nozzle
adjustment tool to rotate it counter-clockwise.
3.
Remove the nozzle bushing (6) with the nozzle insert (7) from
the nozzle adjustment nut by using the nozzle installation tool to
rotate it counter-clockwise.