8.
Quality
Control
2008-07
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
137
Material/Reagents:
1 nm Europium (B119-100, HVD Live Sciences)
Enhancement Solution (=Blank) (1244-105, HVD Live Sciences)
1 Greiner 96-well plate white
200 µl P tips
Calculation of Detection Limit (Sensitivity):
See 8.3.1 Fluorescence Top: Sensitivity.
Precision
Perform the following measurement to determine the precision/reproducibility:
Measurement Parameters:
Parameter Setting
Reading Mode
Fluorescence Top
Exc Wavelength
infinite
®
F200
: 340 nm (35)
infinite
®
M200
: 340 nm
Em Wavelength
infinite
®
F200
: 612 nm (10 nm)
infinite
®
M200
: 617 nm
Number of flashes
25
Integration Time
400
Lag Time
100
Time between Move and Flash
0
Gain Optimal
Plate Type
GRE96fw
Part of the plate
A1
Kinetic 20
Cycles
Interval Time
Minimal
Plate Layout:
See 8.3.3 Time Resolved Fluorescence Precision.
Calculation of Precision: