7. Instrument Features
120
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
2008-07
7.5 Absorbance
Parameters Instrument
Type
Characteristics
Wavelength Range
infinite
®
F200
230 – 1000 nm
infinite
®
M200
230 – 1000 nm
no filter necessary, selection in
1 nm steps possible
Measurement Range Both
0 – 3 OD
The following specifications are valid for the wavelength range from
300 – 700 nm for
infinite
®
200
series
.
Plate type (number of wells)
96
Accuracy 0 – 2 OD
<
±
(1 % + 10 mOD)
Accuracy 2 – 3 OD
<
±
2.5 %
Baseline Flatness
±
10 mOD (1 sigma)
M200
Wavelength Accuracy
≤
± 1.5 nm
λ
> 315 nm;
≤
± 0.8 nm
λ
≤
315 nm
F200
Wavelength Accuracy
Dependent on filters used
The specifications are valid for measurements performed with 25 flashes (reads)
per well.