![tecan infinite 200 Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/tecan/infinite-200/infinite-200_instructions-for-use-manual_828982036.webp)
2. General Description
36
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
2008-07
2.3.6
Injector Reagent Compatibility
The injector system of the infinite
®
F/M200 consist of the following materials:
•
Teflon (PTFE): Tubing, valve plug, seal
•
KelF: Valve body
•
SC05: Injector needles
Please refer to the following list for reagent compatibility. The column ‘Rating’
indicates the compatibility with the reagents listed in the ‘Chemical’ column.
Rating ‘A’ indicates a good compatibility with the injector system. Chemicals with
a rating ‘D’ must not be used with the infinite
®
injectors. They will severely
damage the injector system.
Chemical
Rating
Acetic Acid < 60%
A
Acetonytrile A
Butyl Amine D
Carbon Tetrachloride (dry) D
Chloroform A
Diethyl Ether D
Dimethyl Formamide
A
Ethanolamine D
Ethanol
A
Ethylene Diamine D
Furfural D
Hexane A
Hydrofluoric Acid D
Methanol (Methyl Alcohol)
A
Monoethanolamine D
Potassium Hydroxide (Caustic Potash) D
Potassium Hypochlorite D
Sodium Hydroxide D
Sodium Hypochlorite D
Concentrated Sulfuric Acid D
Diluted Sulfuric Acid (Concentration
≤
1 N)
A
Tetrahydrofuran A
Water, Deionized
A
Water, Distilled
A
Water, Fresh
A