8.
Quality
Control
2008-07
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
143
8.3.8
Absorbance Baseline Flatness (
infinite
®
M200
)
Perform the following measurement to determine the baseline-flatness:
Measurement Parameters:
Parameter Setting
Reading Mode
Absorbance
Measurement Wavelength from/to
300 - 700 nm
Number of flashes
25
Settle Time
0
Plate Type
GRE96ft
Part of the Plate
A1
Plate Layout:
No plate is necessary for measurement – the plate carrier has to be empty for this
measurement.
Material/Reagents:
No material or reagents necessary for this test.
Calculation of Baseline Flatness:
Calculate the standard deviation.