6. Operating the infinite® 200
94
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
2008-07
6.2 General Operating Features
The
infinite
®
200
has some general behavior and options, which are independent
from a particularly selected measurement technique.
6.2.1
Instrument Start Up
Before the instrument is switched ON, check if the USB interface cable is
connected.
Instrument Power On
When switching ON the instrument no initialization steps are performed.
Connect to Instrument
When the software connects to the instrument, communication is established
between the instrument and the user interface.
The following steps are performed:
•
Initialization of OS filter wheels (M200 only)
•
Initialization of luminescence filter wheel
•
Initialization of z-transport of luminescence optics
•
Initialization of plate transport
•
The plate transport is not moved out automatically.
The current versions of firmware and software are displayed.
The instrument is ready to be operated.
6.2.2
Finish a Measurement Session
Disconnect from Instrument
When disconnecting, communication between the instrument and the PC is
terminated.
Note
Remove the microplate and/or cuvette before disconnecting
the instrument from the computer.
Instrument Shut Down
Upon shut down, the instrument activity is stopped immediately. Normally, you
should disconnect before shut down. In the rare case of an unexpected hardware
error, immediate instrument shut down will reduce the risk of possible damage.