![tecan infinite 200 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/tecan/infinite-200/infinite-200_instructions-for-use-manual_828982058.webp)
4. Defining Filter Slides (infinite® F200 only)
58
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
2008-07
4.3 Installing a Custom Filter
When installing a new filter use the filter assembly tool included in the
accessories case. For installing the polarizers use the soft tweezers (plastic).
4.3.1
Removing a Filter
1. Align the filter assembly tool with the notch of the stop-ring. Turn the tool
and remove the stop-ring by pulling it out of the filter slot.
2. The filter will slide out of the filter slot when the filter carrier is turned over.
Do not use the filter assembly tool to remove filters.
4.3.2
Mounting a Custom Filter
A new filter (and polarizer) must be inserted into the slide as shown below.
Note
Make sure that the filters are inserted correctly (see Filter Types). To
ensure proper function, do not reuse the stop-rings more than 5 times.
STOP
Caution
Take care to insert the polarizers and the filters into the filter slide when
working with fluorescence polarization.
Filter
Filter slide
Stop-ring
Stop-ring
Polarizer