![tecan infinite 200 Скачать руководство пользователя страница 129](http://html1.mh-extra.com/html/tecan/infinite-200/infinite-200_instructions-for-use-manual_828982129.webp)
8.
Quality
Control
2008-07
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
129
8.3 Specifications - Test Instructions
8.3.1
Fluorescence Top
For the
infinite
®
200
with the option ‘Fluorescence Top’, the following tests can
be performed to prove the specifications:
•
Sensitivity
•
Uniformity
•
Precision
These test instructions are valid for the
infinite
®
200
:
•
infinite
®
F200
•
infinite
®
M200
•
Standard version
•
Spectrally enhanced version
Sensitivity
Perform the following measurement to determine the detection limit for
Fluorescein:
Measurement Parameters:
Parameter Setting
Reading Mode
Fluorescence Top
Exc Wavelength
infinite
®
F200
: 485 (20 nm)
infinite
®
M200
: 485 nm
Em Wavelength
infinite
®
F200
: 535 (25 nm)
infinite
®
M200
: 535 nm
Number of flashes
25
Integration Time
40
Settle Time
0
Gain Optimal
Plate Type
GRE96fb