Rosemount TankRadar Rex
Chapter 1 Checks and Adjustments
1-24
Commissioning Manual
308011En, Ed.2/Rev.B
June 2008
1.14.6
Software
•
Write protection (on/off)
•
If you use the DB loader, note the version. Normally, the DB
loader is only used in connection with software upgrading.
•
Note which version of the DAU program that is used.
1.15 DAU Configuration
1.15.1
Tank Information
•
Tank name
•
Check that the Unit ID corresponds to the DAU installed at the cur-
rent tank
•
Check that the address corresponds to the address specified in
TankMaster of the current tank.
1.15.2
Temperature Check
•
Check that each temperature sensor outputs a correct value.
•
Note the average temperature.
1.15.3
Current inputs
•
Note the number of current inputs used, and check the measured
values.
1.15.4
Display
•
Check that no
error codes
are displayed.