Rosemount TankRadar Rex
Chapter 1 Checks and Adjustments
1-16
Commissioning Manual
308011En, Ed.2/Rev.B
June 2008
1.12.2
Visual check
•
Note if the pipe and pins are inspected by a service engineer.
•
Check that a deflection plate with calibration ring, and a verifica-
tion pin are mounted
.
•
Check that the verification pin is aligned towards the flangemark,
see Figure 1-15.
•
Check that the waveguide, housing and foam plug are installed
according to the instructions in the
Installation Manual, chapter 10
.
1.12.3
Still-Pipe
•
Note the number of holes, the distance between the holes (unit
meter) and the diameter (unit mm).
•
Note the pipe inner diameter.
•
Check that the pipe cone fits tightly to the still-pipe.
Figure 1-15 Align the verification pins.
Note!
The topmost hole/slot in the still-pipe must be placed above maximum
product level.
Align!
Flange mark indicating
the direction of the
Verification pins
Verification pin
Still-pipe