Calibration
------------------------------------------------------------------------------------
179
Calibrating Meters
Preparation
The following measuring devices are required for calibration.
1.Digital Multimeter for Voltage/Current Measurement- 1 set
2. Shunt for Current Measurement – 1 set
No. of Display Digits : 5 1/2 or more
Display Accuracy : 0.005% or more
(Possesses a measuring range from 1mV~100V)
50A/50mV 0.1% accuracy (400LN/LAN)
100A/50mV 0.1% accuracy (800LN/LAN)
200A/50mV 0.1% accuracy (1600LN/LAN)
Voltmeter Offset/Full-Scale Calibration
1. Disconnect the load wires and the remote sensing
wires. Connect the digital multimeter to the output
terminal.
Turn the POWER switch ON and OUTPUT OFF in
advance.
2. Press the PRESET key ① and set to “0.01V” with
the VOLTAGE dial ② and “1.00A” with the
CURRENT dial ③.
(See Operating as Constant Voltage Power Supply
under Basic Operations on how.)
3.Press the PRESET key again and then the OUTPUT
key to set the output to ON.
4. Press the FUNC. key and change the top Number
Display to Settings Item 24 with the VOLTAGE dial
to display the voltage measurement value at the
bottom.
5. Adjust the measurement value displayed in the
bottom Number Display to the same value as
displayed on the digital multimeter (used for output
voltage measurement) with the CURRENT dial.
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
Calibration Circuit
Digital
Multimeter
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
①
②
Note: Bottom Number Display
is displayed in volts(V).
+
-
V
Power Supply
+
-
Before performing this operation, ensure “Output Voltage Setting Offset/Full-scale Calibration”
is carried out.
Caution
Set FUNCTION Settings Item 13, Linearity compensation
Function, to 0 (0=OFF) beforehand.
Caution
①
②
③
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
D ISPL AY
A
B
C
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
ZX-S-400 LN
Содержание ZX-S-1600LAN
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Страница 13: ...About this Product 10 This is blank page...
Страница 63: ...Useful Function 60 This is blank page...
Страница 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Страница 157: ...Special Loads 154 This is blank page...
Страница 199: ...Specifications 196 This is blank page...
Страница 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Страница 201: ...Appendix 198 ZX S 800LN 800LAN...
Страница 202: ...Appendix 199 ZX S 1600LN 1600LAN...