
Calibration
------------------------------------------------------------------------------------
159
Calibrating Output Settings
Preparation
The following measuring devices are required for calibration.
1.
Digital Multimeter for Voltage/Current
Measurement- 1 set
No. of Display Digits : 5 1/2 or more
Display Accuracy : 0.005% or more
(Possesses a measuring range from 1mV~100V)
2.Shunt Resistor for Current Measurement- 1 set
50A/50mV 0.1% accuracy (400LN/LAN)
100A/50mV 0.1% accuracy (800LN/LAN)
200A/50mV 0.1% accuracy (1600LN/LAN)
Offset/Full Scale Calibration of Output Voltage Settings
1. Disconnect the load wires and remote sensing wires.
Connect the digital multimeter to the output terminal.
4.Press the FUNC. key and change the top Number
Display to Settings Item 20 with the VOLTAGE
dial.
In advance, turn POWER switch ON and OUTPUT OFF.
2.Press the PRESET key ① and set to “0.01V” with
the VOLTAGE dial ② and “1.00A” with the
CURRENT dial ③.
(See Operating as Constant Voltage Power Supply
under Basic Operations on how.)
3.Press the PRESET key again and then press the
OUTPUT key to set output to ON.
5.Adjust so that the digital multimeter (used for
output voltage measurement) displays “0.010V”
with the CURRENT dial.
6 . Press the FUNC. key to complete offset
calibration of the output voltage settings.
FUNC. lamp will light and the display at bottom Number Display
will disappear.
MEMO
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Calibration Circuit
Digital
Multimeter
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
①
②
+
-
V
Power Supply
+
-
①
②
③
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
D ISPL AY
A
B
C
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
ZX-S-400 LN
Содержание ZX-S-1600LAN
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Страница 13: ...About this Product 10 This is blank page...
Страница 63: ...Useful Function 60 This is blank page...
Страница 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Страница 157: ...Special Loads 154 This is blank page...
Страница 199: ...Specifications 196 This is blank page...
Страница 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Страница 201: ...Appendix 198 ZX S 800LN 800LAN...
Страница 202: ...Appendix 199 ZX S 1600LN 1600LAN...