Calibration
------------------------------------------------------------------------------------
161
Offset/Full-Scale Calibration of Output Current Settings
1.Perform load with only the shunt resistor.
3.Press the PRESET key again and then press the
Disconnect the load wires and remote sensing wires.
OUTPUT key to set output to ON.
Connect the shunt to the output terminal.
Connect the digital multimeter to the voltage terminal
located on the shunt resistor.
Turn the POWER switch ON and OUTPUT OFF in
advance.
4.Press the FUNC. key and change the top Number
Display to Settings Item 22 with the VOLTAGE dial.
2. Press the PRESET key ①. Set the CC set value to
“0.1A” (※) with the CURRENT dial ③ and set the
CV set value to “1.00V” with the Voltage dial ②.
※ Set the CC set value to “0.01A” for 400LN/LAN,
800LN/LAN and “0.1A” for 1600LN/LAN power supplies.
(
See Operating as Constant Current Power
Supply under Basic Operations on how.
)
5 . Adjust so that the digital multimeter (used for
current measurement) displays to “0.010mV” (For
400LN/LAN use 10mA) (※) with the CURRENT dial.
※For 800LN/LAN use “0.005mV”(equivalent to 10mA)
For 1600LN/LAN use “0.025mV”(equivalent to 100mA)
Set FUNCTION Settings Item 1, Sink Function, to 1 (1=ON)
and Settings Item 13, Linearity compensation Function, to
0(0=OFF)beforehand.
Caution
FUNC. lamp will light and the display on the bottom Number
Display will light out.
MEMO
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
①
②
Note:
Current Flow is 40A for
(400LN/LAN)
800LN/LAN is 80A
1600LN/LAN is 160A
Use wires which will not
produce a voltage drop
below 0.8V.
+
-
Digital
Multimeter
Power Supply
V
Shunt
-
+
Calibration Circuit
①
②
③
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
D ISPL AY
A
B
C
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
ZX-S-400 LN
Содержание ZX-S-1600LAN
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Страница 13: ...About this Product 10 This is blank page...
Страница 63: ...Useful Function 60 This is blank page...
Страница 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Страница 157: ...Special Loads 154 This is blank page...
Страница 199: ...Specifications 196 This is blank page...
Страница 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Страница 201: ...Appendix 198 ZX S 800LN 800LAN...
Страница 202: ...Appendix 199 ZX S 1600LN 1600LAN...