SynCardia Systems, LLC
Companion 2 Driver System Operator Manual
Page 94 of 132
14.5. Notifications
The Driver provides audible and visual notifications (
Table 14-3
) that do
not represent alarm conditions. These notifications are intended to
improve communication of events or actions.
Table 14-3 – Notifications
Event
Visual Indicators
Audible Signal
Corrective Action
External Power
Source
Connection
External Power
Indicator: Solid Green
Power Status Window:
Plug Icon
Two rising tones
in succession
No action required
External Power
Source
Disconnect
External Power
Indicator: Yellow
Power Status Window:
Remaining battery
support time
Two falling tones
in succession
No action required unless other
power related alarm conditions
exist, e.g., low External Battery
power
External Battery in
Use
Numerical display of the
estimated remaining
battery support time.
No audible signal
No action required
Emergency
Battery in Use
Power Status Window:
Icon overlays the
External Battery Icon
Numerical display of the
estimated remaining
battery support time
10 beeps every
2.5 seconds
Connect Driver to an external
power source (wall outlet).
Replace External Batteries with
fully charged Batteries.
If alarm continues, replace Driver
with backup Driver.
Emergency
Battery Low
External Power
Indicator: Solid Green
Emergency Battery
Indicator: Flashing Red
Power Status Window:
Plug Icon
No audible signal
No action required. Emergency
Battery is charging. Emergency
Battery Low Alarm will clear
when charged to adequate level.
OR Mode
Notification
Display shall indicate it
is in OR Mode
Two beep audible
reminder signal
every 2 hours
No action required if the device is
in use during a surgical
procedure, otherwise change to
I.C.U. Mode or Ambulatory Mode
to activate audible alarms
Audible Alarms
Disabled in O.R.
Mode Reminder
Solid “X” over the Alarm
Mute Button
Two beep audible
reminder signal
every 2 hours
No action required if the device is
in use during a surgical
procedure, otherwise change to
I.C.U. Mode or Ambulatory Mode
to activate audible alarms
Содержание Companion 2 Driver System
Страница 2: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 2 of 132...
Страница 4: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 4 of 132...
Страница 8: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 8 of 132...
Страница 14: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 14 of 132...
Страница 16: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 16 of 132...
Страница 18: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 18 of 132...
Страница 32: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 32 of 132...
Страница 56: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 56 of 132...
Страница 66: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 66 of 132...
Страница 96: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 96 of 132...
Страница 118: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 118 of 132...
Страница 130: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 130 of 132...
Страница 132: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 132 of 132...