![SynCardia Companion 2 Driver System Скачать руководство пользователя страница 87](http://html1.mh-extra.com/html/syncardia/companion-2-driver-system/companion-2-driver-system_operators-manual_1420565087.webp)
SynCardia Systems, LLC
Companion 2 Driver System Operator Manual
Page 87 of 132
Alarm
Message
Visual Indicators
Audible Signal
Corrective Action
Left Fill
Volume High
NOTE:
this alarm only
applies to Driver use with
the 70cc TAH-t (this alarm
is not present with the
50cc TAH-t)
Left fill volume:
Background is yellow.
Alarm Bar: Flashing
Yellow
Alarm Indicator LED:
Flashing Yellow
3 beeps every 30
seconds
70-80 dBA
Verify Driver settings are achieving full
eject/partial fill conditions from Driver
displayed wave forms.
Assess patient conditions that may
contribute to a Left Fill Volume High
alarm.
If alarm persists, certified clinical staff
may make adjustment to device rate or
vacuum.
Right Fill
Volume High
NOTE:
this alarm only
applies to Driver use with
the 70cc TAH-t (this alarm
is not present with the
50cc TAH-t)
Right fill volume:
Background is yellow.
Alarm Bar: Flashing
Yellow
Alarm Indicator LED:
Flashing Yellow
3 beeps every 30
seconds
70-80 dBA
Verify Driver settings are achieving full
eject/partial fill conditions from Driver
displayed wave forms.
Assess patient conditions that may
contribute to a Right Fill Volume High
alarm.
If alarm persists, certified clinical staff
may make adjustment to device rate or
vacuum.
Incorrect
Percent
Systole
Alarm Bar: Yellow
Alarm Indicator LED:
Yellow
Percent Systole
Background is: Yellow
2 beeps every 3½
minutes
70-80 dBA
Replace Driver with a SynCardia
approved backup Driver.
Notify SynCardia at
1-866-771-9437
Configuration
Data Lost
Alarm Bar: Yellow
Alarm Indicator LED:
Yellow
2 beeps every 3½
minutes
70-80 dBA
Replace Driver with a SynCardia
approved backup Driver.
Notify SynCardia at
1-866-771-9437
Filter
Maintenance
Required
Alarm Bar: Yellow
Alarm Indicator LED:
Yellow
No audible signal.
Using the finger access groove,
replace/clean filter.
Driver
Requires
Service
Alarm Bar: Yellow
Alarm Indicator LED:
Yellow
2 beeps every 3½
minutes
70-80 dBA
Replace Driver with a SynCardia
approved backup Driver.
Notify SynCardia at
1-866-771-9437
Содержание Companion 2 Driver System
Страница 2: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 2 of 132...
Страница 4: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 4 of 132...
Страница 8: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 8 of 132...
Страница 14: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 14 of 132...
Страница 16: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 16 of 132...
Страница 18: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 18 of 132...
Страница 32: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 32 of 132...
Страница 56: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 56 of 132...
Страница 66: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 66 of 132...
Страница 96: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 96 of 132...
Страница 118: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 118 of 132...
Страница 130: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 130 of 132...
Страница 132: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 132 of 132...