![SynCardia Companion 2 Driver System Скачать руководство пользователя страница 86](http://html1.mh-extra.com/html/syncardia/companion-2-driver-system/companion-2-driver-system_operators-manual_1420565086.webp)
SynCardia Systems, LLC
Companion 2 Driver System Operator Manual
Page 86 of 132
Alarm
Message
Visual Indicators
Audible Signal
Corrective Action
Computer
Malfunction
Alarm Bar: Red
Alarm Indicator LED: Red
10 beeps every 5
seconds
70-80 dBA
Computer Malfunctions are intended to
notify the user of software
communication errors. This alarm may
recover automatically; however, if the
alarm persists, the Driver should be
replaced with a SynCardia approved
backup Driver.
Notify SynCardia at
1-866-771-9437
Left Vacuum
Incorrect
Alarm Bar: Yellow
Alarm Indicator LED:
Yellow
2 beeps every 3½
minutes
70-80 dBA
Verify Driver Left Vacuum Incorrect
alarm settings are appropriate (
Section
14.4
).
If alarm persists, replace Driver with a
SynCardia approved backup Driver.
Notify SynCardia at
1-866-771-9437
Right
Vacuum
Incorrect
Alarm Bar: Yellow
Alarm Indicator LED:
Yellow
2 beeps every 3½
minutes
70-80 dBA
Verify Driver Right Vacuum Incorrect
alarm settings are appropriate (
Section
14.4
).
If alarm persists, replace Driver with a
SynCardia approved backup Driver.
Notify SynCardia at
1-866-771-9437
External
Battery Error
Alarm Bar: Yellow
Alarm Indicator LED:
Yellow
Battery Indicator: Yellow
2 beeps every 3½
minutes
70-80 dBA
Verify Driver is connected to an
external power source (wall outlet) and
External Power indicators are
displayed.
Replace External Batteries with
charged Batteries and verify Batteries
indicate a charge on battery fuel
gauge.
If alarm persists, replace Driver with a
SynCardia approved backup Driver.
Notify SynCardia at
1-866-771-9437
Low External
Battery
Alarm Bar: Yellow
Alarm Indicator LED:
Yellow
Battery Indicator: Yellow
2 beeps every 3½
minutes
70-80 dBA
Verify Driver is connected to an
external power source (wall outlet) and
External Power indicators are
displayed.
Replace External Batteries with
charged Batteries and verify Batteries
indicate a charge on battery fuel
gauge.
If alarm persists, replace Driver with a
SynCardia approved backup Driver.
Notify SynCardia at
1-866-771-9437
Содержание Companion 2 Driver System
Страница 2: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 2 of 132...
Страница 4: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 4 of 132...
Страница 8: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 8 of 132...
Страница 14: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 14 of 132...
Страница 16: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 16 of 132...
Страница 18: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 18 of 132...
Страница 32: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 32 of 132...
Страница 56: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 56 of 132...
Страница 66: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 66 of 132...
Страница 96: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 96 of 132...
Страница 118: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 118 of 132...
Страница 130: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 130 of 132...
Страница 132: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 132 of 132...