12.5. Fuses
Fuses are user replaceable.
12.6. Reduction of Explosion Hazard
Do NOT operate the Companion 2 Driver System in the presence of
flammable anesthetics or other flammable gases.
12.7. Maintenance of Patient Safety
Verify that all pneumatic and electrical connections are in place, secure,
and connected to the Driver in use before attempting to operate the
System.
12.8. Environment
Keep the Driver dry. Protect it from showers, baths, rain, and liquid spillage.
DO NOT operate the Driver beyond an ambient pressure of 525 mmHg to 795
mmHg (700 to 1060 mBar), or at altitudes less than 2000 ft. below sea level or
above an altitude of 8,000 ft. above sea level. Driver displayed Cardiac Output
is accurate to within 10% throughout this range. Above 4000 ft Driver
displayed Cardiac Output is accurate to within 15% for hypertensive patient
conditions.
12.9. Removal of Key During Operation
Remove the key from the Key Switch when the Driver is in operation. The
key cannot be removed when the Driver is switched OFF.
12.10. Driver Storage Requirements
Always store Companion 2 Drivers in a Hospital Cart or Caddy that is
connected to external (wall) power. Storing the Driver without external
(wall) power may cause the Emergency Battery to deplete and may render
the Driver unusable for patient support.
The Driver should always be connected to an active wall outlet at any time
when the patient is not ambulating in order to conserve battery power.
Batteries in the Driver are charged when wall power is applied. The Driver
does not have to be set to the ON position to charge Batteries. Additional
Batteries may be charged in the backup Driver. Do not store the
Companion External Batteries in a Companion 2 Driver that is not
connected to external (wall) power. This may render the External Batteries
unusable. External Batteries can be stored separately from the Driver or in
a Companion 2 Driver docked into a Hospital Cart or Caddy that is
connected to external (wall) power.
Содержание Companion 2 Driver System
Страница 2: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 2 of 132...
Страница 4: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 4 of 132...
Страница 8: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 8 of 132...
Страница 14: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 14 of 132...
Страница 16: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 16 of 132...
Страница 18: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 18 of 132...
Страница 32: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 32 of 132...
Страница 56: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 56 of 132...
Страница 66: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 66 of 132...
Страница 96: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 96 of 132...
Страница 118: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 118 of 132...
Страница 130: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 130 of 132...
Страница 132: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 132 of 132...