70
4.3 USE
4.3.1 Precautions
To ensure optimum measurement precision avoid all lateral pressure when presenting the probe contact to the objet
to be measured. Ideally, a mechanical retracting lifter should be used.
Carefully clamp the fixing bearing of the probe in the holder. Fixing too tight can influence the measurement.
Avoid any impact on the probe spindle.
4.3.2 Changing the contact point
The probe spindle has an ISO M 2.5 thread in the end allowing replacement of the contact points. When changing
the contact point, the probe spindle should be in the outmost position.
4.4 MAINTENANCE
This has been reduced to a simple operation. When the measuring spindle no longer slides with complete ease
And precision, clean it with a dust free cloth and lightly lubricate with a fine oil.
4.4.1 Replacing the connection cable, for P10, P25 and P50 :
unscrew the cap (
a
)
pull out the cable
plug in the new cable, locate the pin (
b
) of the cable sleeve (
d
) in the slot (
c
).
Screw on the cap (
a
)
Notes : use connection cable for the corresponding probe (
P10-P25-P50
).
For the
P2/P5
:the cable is connected to the probe by means of a sealed connector.