29
1.6.11 SETUP KEY
Allows input of parameters for the following functions : Stat, Class, Tol, Clear or Enter, Min/Max, SetUp, Preset,
Store, Print, Scan et Channel. The input of parameters is explained with each corresponding function key.
Input of general parameters of unit :
1/ Press
Setup
key .2/ then
Setup
key. Following will be displayed :
To exit this menu or the followings without changes,
press
Clear
key or
Enter
key.
1.6.11.1 External contact 1 function :
menu 1 allows you to define the external contact function (in general the foot pedal supplied with the unit).
When the unit is ready for measurements, it is thus possible to inhibit the keyboard and use only the foot-
pedal.
1
/: Transmission of values through the Centronics or RS-232-C output according to the parameters specified for
the Print function. Each time the foot pedal is pressed, the measurement value is transmitted.
This choice may be combined with the following other functions :
2 : display hold/store. The display is held as long as the foot pedal is pressed, when foot pedal is released, the
measured value is sent. The function 5 (change channels),6 (D110 command), 7 (scanning) can be combined as
well.
3 : Min/Max initializationMin/Max. The display is held as long as the foot pedal is pressed, when foot pedal is
released, the measured value is sent. The functions 2 (hold/store), 5 (change channels) can be combined as well.
4 : display preset. Alternately, pressing the pedal presets the display, pressing it again sends the measured value.
5 : change channel. The same action on the foot pedal sends the measured value and then changes to the next
channel.
6 : control the lifting of probes with
D110
unit. The probes are lifted as long as foot pedal is pressed. When foot
pedal is released, the probes will go down and the measured value is sent. A delay can be introduced between
descent and transmission of value (value 1 to 36000 = 1 sec to 10 hours). Function 7 (start scanning) can be
combined as well. The first time the pedal is pressed the probes will be lifted, the second time they will go down,
scanning starts (after the delay, if any ) the value is sent.
7 : scanning starting. Scanning is carried out first, followed by data transmission. A delay between the release of the
foot pedal and the scanning can be entered.
If
a combination of the second or 3
rd
function is not required, simply press
Enter
key.