38
1.6.14.2 Peripheral choice, the following menu will appear :
The following will be displayed :
These different choices modify the output format for the Centronics/RS232
outputs.
Remote command of the unit will not be modified.
80 and 40 column formats allow the header to be printed following the user’s
parameters.
2.2.1/ Connection to a computer. This is the most simple transmission format allowing easy processing of values.
The measurement is transmitted as it appears on the display :
a/ In mm
:
SIGN 10
2
10
1
10
0
DP 10
-1
10
-2
10
-3
10
-4
CR LF
or 10
3
SIGN = space if positive sign DP = decimal point LF only if requested
10
2
et 10
1
= space3 if zero 10
-4
, 10
-3
and 10
-2
only with resp. resolutions 0.1 µm, 1 µm et 10 µm.
The sign always immediately precedes the 1
st
digit.
b/ In inch
:
SIGN 10
1
10
0
DP 10
-1
10
-2
10
-3
10
-4
10
-5
CR LF
10
1
= space if zero 10
-5
, 10
-4
et 10
-3
only with resp. resolutions 0.00001, 0.0001et 0.001 in
c/ In scanning mode the values are transmitted as follows
:
5. < = number of measured work piece followed by< (reject), = (good) or > (rework) if TOL active.
10.002 = measured value of channel 1
20.002
12.003
7.999 to channel n..
+
CR
(+
LF
) extra to indicate end of transmission of measurements.
d/
Stored values
are transmitted from first to last. Between each measurement, a
CR
(+
LF
) is transmitted.
e/
Statistical values are transmitted as follows
:
CANAL 1 = Channel No. for which statistics are transmitted.
N 28 the statistical values are transmitted in the same order as they appear on the screen
Xmax 10.004 (
refer to section 1.6.9
). The histogram is not transmitted.
Xmin 9.998
R 0.006
Xbar 10.000
Sigma 0.0017
s 0.0017
- 3 S 9.995 This value and the following are only transmitted if tolerances are activated.
+3 S 10.005
- NG 7
+ NG 6
%Def 46.428
Cp 0.2986
Cpk 0.2450
Cm 0.2932
Cmk 0.2290