33
Example : The probe is used for measuring a camshaft. The unit displays the maximum value. The camshaft is
rotated and the displayed value is frozen on the maximum value recorded. The preset value is entered, for example
10,000 mm.
Min/Max registers are initialized by pressing
Clear
key.
A new measurement of the camshaft will now display value of 10.000 mm.
It is possible to combine A +/- B mode with Min/Max mode (
refer to section 1.6.3
) Channel A or channel B both
channels can be set in one of the Min/Max modes described above. If channel A is in Min/Max mode, both
measurements cannot be taken simultaneously, but only one after the other. The channel A measurements is taken
first, then by pressing ENTER Key, the channel B measurement will follow (the channel in use is indicated on the
display in reverse video). If only channel B is in Min/Max mode, the unit switches from A to B automatically.
The
Clear
key reinitializes Min/Max registers in case of error either in channel A or B (switching from channel A to B
or the opposite will automatically initialize Min/Max registers for the new active channel).
The scanning mode (refer to section 1.6.13) is also compatible with A +/- B mode combined with Min/Max mode. In
this case, scanning stops on the channel in Min/Max mode and waits for measurement. The measured value is then
displayed. Scanning may be continued by pressing the ENTER key or the foot pedal (depending on what function
has been assigned to it).