37
1.6.14 PRINT KEY
Printing of values through :
Centronics
output, identical to parallel port of PC (LPT1 or 2). This output does not need to be set up and allows
connection of practically all printers on market. Different printing formats are generated by unit
D100S
: 80, 40 or 15
columns.
Output RS-232-C, in accordance with transmission parameters selected for this function.
A
P
in reverse video is displayed on the top right corner of display during RS 232-C / Centronics transmission. If the
channel has no installed probe (NO PROBE on display), 999.9 is transmitted.
It is also possible to configure the external contact (foot pedal) for the RS-232-C/Centronics output (refer to section
1.6.11)
Setting RS 232 transmission parameters and Centronics/RS-232-C output format :
1/ press
Setup
key 2/ then
key the following will be displayed :
The 1
st
menu sets RS 232 C transmission parameters.
The 2
nd
menu select the Centronics/RS-232-C output format according to
which peripheral is being used.
1.6.14.1 RS232 input/output parameters :
Default parameters (after a reset) :
4800 bps, 7 bits, even parity, CR
.
2.1.1/ transmission speed : 300, 600, 1200, 2400, 4800, 9600 et 19200 bauds/sec.
By pressing the
Clear
key the preceding menu will be displayed.
By pressing the
Enter
key the next menu will be displayed (without changing previous choice).
2.1.2/ word length : on
7
or
8
bits.
2.1.3/ parity control :
No
parity,
even
or
odd
.
2.1.4/ end characters :
CR
(Carriage Return) or CR + LF (Carriage Line Feed) or LF only. A printer with
auto LF mode needs only CR. If CR + LF is sent in this case, a supplementary empty line is printed at each
carriage return. With a connection to an PC or compatible select CR only.