Setting Menu
Page 16/28
Programming instructions of the
SYCLOPE INDIG’O
®
Messages and signification after calibrating.
Message
Signification
Action
Fig.
Slope OK
Calibration correct. Slope changed
Calibration done
1
Offset OK
Calibration correct. Offset changed
Calibration done
-
Slope LIMIT
Check the probe
The calculated slope after calibration is
out of normal value. (Calculated
slope<> 33…..250% of standard slope)
Calibration done. Display of
the pictogram key of
maintenance on the
permanent screen
2+6
Offset LIMIT
Check the probe
The calculated slope after calibration is
out of normal value.
Calibration done. Display of
the pictogram key of
maintenance on the
permanent screen
-
Slope error
p=25…300%
(Version 4-20 mA)
The calculated slope after calibration is
out of acceptable values between 25
and 300%
Calibration refused
3
Offset error
Delta (max) = ±90 mV
(Version pH)
The calculated offset after calibration is
out of acceptable values
Calibration refused
Calibration error
Delta (max) = ±70 mV
(Version ORP)
The calculated offset after calibration is
out of acceptable values
Calibration refused
Can’t be calibrated
during polarization
Impossible calibration during
polarization
Calibration refused
4
Reset with factory
calibration
Reset calibration (slope and offset)
values
Back to factory value
5
Control Alarms
Calibration
Sensor: 6,506 mA
Raw value:
2,08 mg/l
2,09
mg/l
Buffer:
Slope : 1,19 mA/mg/l
@Standard :
100
%
Slope OK
Control Alarms
Calibration
Sensor: 6,506 mA
Raw value:
2,08 mg/l
2,09
mg/l
Buffer:
@Standard :
100
%
Slope : 1,19 mA/mg/l
Slope LIMIT
Check the probe
Control Alarms
Calibration
Sensor: 6,506 mA
Raw value:
2,08 mg/l
2,09
mg/l
Buffer:
@Standard :
100
%
Slope : 1,19 mA/mg/l
Slope error
p=25…300%
Control Alarms
Calibration
Sensor: 6,506 mA
Raw value:
2,08 mg/l
2,09
mg/l
Buffer:
@Standard :
100
%
Slope : 1,19 mA/mg/l
Reset with
factory calibration
Free Cl
mg/l
2,09
REG
ALR
Dosage : 58%
Set point : 2,00 mg/l
Output : 4,187 mA
Control Alarms
Calibration
Sensor: 6,506 mA
Raw value:
2,08 mg/l
2,09
mg/l
Buffer:
@Standard :
100
%
Slope : 1,19 mA/mg/l
Can’t be calibrated
during polarization
1
2
3
4
5
6