Configuration Menu
Page 22/28
Programming instructions of the
SYCLOPE INDIG’O
®
3)
Range for 4-20 mA analog output : Page “Analog output”
This page allow you to set the range of the analog output
I
OUT1
:
Range : define the range of the analog output. This ouput allow to transfer the measuring value
to a PLC.
Name
Signification
Setting range
Default value
Range
Working range of the output
0-20 mA or 4-20 mA
4-20 mA
4)
Flow switch input : Page “Flow switch”
This page allow you to define the type of the input K
1pause
. When the condition is not checked, the
controller put in stand-by the control and display a pictogram on the main screen.
Type : define the type of the input NO or NC
NO : Normally open
NC : Normally Close
Name
Signification
Setting range
Default value
Type
Type of the input K
1pause
NO or NC
NC
5)
Polarization time : Page “Sensor”
This page allow you to define the polarization time of the sensor. After the power supply or after a
default on the sensor, the controller wait during the time set before controlling. During this time, a
pictogram is displayed on the main screen:
Analog output
ontrol relay
4-20 mA
0-20 mA
Range :
Flow switch
Analog output
NC
NO
Type :
Free Cl
1,09
REG
ALR
mg/l
Setpoint : 2,00 mg/l
Dosage : 0,00%
Output : 5,750 mA