12
TRIO MP-235 THREE-AXIS MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.23 (20190130) (FW V2.2 & 2.3)
To remove it, first loosen the four hex screws that secure the manipulator to the pegs in the
adapter plate. The rear pair is in a similar location in the back of the manipulator.
Once the
locking screws are sufficiently loosened, lift the TRIO MP-235/M upwards from the adapter
plate.
Before attaching the adapter plate to the TRIO MP-235/M, you need to decide where to
position the manipulator on your stand/platform. The stand can be any flat surface carrying
¼-20, 10-32, or M6 holes on one-inch centers (such as a Sutter Instrument MT-series stand
or MD series platform).
Examine the space of the platform onto which installation is to take place. Attach the control
cable to TRIO MP-235/M and move the entire unit around on the platform until the precise
desired position is determined.
A small bag containing the necessary hardware to attach the
TRIO MP-235 to the stand is included.
Figure 2-2. Mounting the TRIO MP-235/M on the Adapter Plate
Once the plate is mounted, align the pegs on top of the plate with the holes in the
manipulator, push the X-axis firmly onto the plate, and re-tighten the locking hex set
screws.
2.2
Headstage Mounting
Sutter IPA headstage, Axon headstages 203B or CV-7, and the Heka EPC-10 headstage have
an integral dovetail that fits directly into the rotary dovetail slide bracket on the TRIO MP-
235/M. The dovetail slide bracket on the TRIO MP-235/M also supports older Axon and Heka
headstages when using the 4’’ dovetail extension.
Rod-mounted headstages and micro tools are accommodated by the use of a rod clamp that
fits into the dovetail (not shown). All the headstage adapters and mounting hardware are
included with the manipulator and are shipped in a zip lock plastic bag.
2.3
Other Accessories
One or more accessories may have been ordered and received for mounting the TRIO MP-
235/M and/or modifying the headstage mount to the manipulator (i.e., rotating base,
microscope stage mount, gantry, dovetail extension, etc.). Setup of these accessories is
normally covered in documentation accompanying the accessory.
Содержание TRIO MP-235