16
TRIO MP-235 THREE-AXIS MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.23 (20190130) (FW V2.2 & 2.3)
2.5.2.7
Switch 8: Reserved (Firmware <= v2.2)
This switch is reserved for future use. The factory default position is OFF (up).
2.5.2.8
Switch 8: ROE Axis Movement Knobs Active – All vs. D Only (Firmware v2.3+)
This switch determines if all axis-movement knobs of the ROE are active or only the D-axis
knob.
Table 2-6. ROE Axis movement knobs active – all vs. D only ( (Firmware v2.3+).
ROE Axis Movement Knobs Active – All vs. D Only
Switch #
All Active (X, Y, Z, & D)
D Only (X, Y, & Z inactive)**
8 OFF
(Up)*
ON (Down)
* Factory default (recommended normal operation setting).
** Use the ON (down) setting for when it is desired to have only the D-axis movement knob
active, while the rest (X, Y, & Z) are inactive.
2.5.2.9
Switch 9: Y-Axis Travel Length (Firmware v2.2+)
This switch informs the ROE/controller as to the travel length of the Y axis in the connected
micromanipulator eletromechanical or stage. The length of the Y axis on the TRIO MP-235/M
electromechanical micromanipulator is 25mm, so Switch 9 should be set to OFF (up) (factory
default for a standard TRIO MP-235 system). For a variant model with 12.5mm of travel in
the Y axis, Switch 9 must be set ON (down).
Table 2-7. Y-Axis travel length (Firmware v2.2+).
Y-Axis Travel Length
Switch #
25mm 12.5mm
**
9 OFF
(Up)*
ON (Down)
* Factory default (recommended normal operation setting).
** Use the ON (down) setting for half-length Y axis (12.5mm).
CAUTION: Always be certain that the position of Switch 9 correctly matches the physical
length of travel of the Y axis on the connected device. Setting the switch to the OFF (up)
position for 25mm could result in equipment damage if the attached device’s Y axis length of
travel is actually less than 25mm.
2.5.2.10
Switch 10: X Axis Travel Length (Firmware v2.2+)
This switch informs the ROE/controller as to the travel length of the X axis in the connected
micromanipulator eletromechanical or stage. The length of the X axis on the TRIO MP-
235/M electromechanical micromanipulator is 25mm, so Switch 10 should be set to OFF (up)
(factory default for a standard TRIO MP-235 system). For a variant model with 12.5mm of
travel in the X axis, Switch 10 must be set ON (down).