Command
Description
Command Syntax
(command underlined)
Sensor response (underlined)
"a" represents the
single-character address
C3 Request
User
Scale, User
Offset, and
Field
Calibration
Offset
aC3!
aD0!
a00003
000 is the time in seconds until the
measurement is ready and 3 is the
number of values that can be collected
asoc
where, s is the user scale and o is the
user offset (psi), and c is the field
calibration offset (psi).
C4 Request
Standards lab
Calibration
Scale and
Offset
aC4!
aD0!
a00002
000 is the time in seconds until the
measurement is ready and 2 is the
number of values that can be collected
aso
where, s is the scale calibration and o is
the offset calibration(psi).
C5 Request
Quadrature
scale factor,
threshold,
step rate, and
operating
mode for the
unit
aC5!
aD0!
a00004
000 is the time in seconds until the
measurement is ready and 04 is the
number of values that can be collected
astrm
where, s is the Quadrature scale, t is the
quadrature threshold, r is the quadrature
step rate in steps per second, and m is
the operating mode of the analog and
quadrature outputs.
C6 Request
Concurrent
Temperature
and Pressure
Measurement
(Version 2.0
and higher)
aC6!
aD0!
attt04
ttt is the time in seconds until the
measurement is ready and 04 is the
number of values that can be collected
atupv
where t is the temperature, u is the
temperature units, p is the pressure, and
v is the pressure units. Use the XUT
command to set the temperature units
and the XUP command to set the
pressure units.