Command
Description
Command Syntax
(command underlined)
ACCUBUBBLE response
(underlined)
"a" represents the
single-character address
M5 Request
Quadrature
scale factor,
threshold,
step rate, and
operating
mode for the
unit
aM5!
aD0!
a0004
000 is the time in seconds until the
measurement is ready and 4 is the
number of values that can be collected
astrm
where, s is the Quadrature scale, t is the
quadrature threshold, r is the quadrature
step rate in steps per second, and m is
the operating mode of the analog and
quadrature outputs.
M6 Request
Temperature
and Pressure
Measurement
(Version 2.0
and higher)
aM6!
aD0!
attt4
ttt is the time in seconds until the
measurement is ready and 4 is the
number of values that can be collected
a
service request
atupv
where t is the temperature, u is the
temperature units, p is the pressure, and
v is the pressure units. Use the XUT
command to set the temperature units
and the XUP command to set the
pressure units.
M7 Request
factory
calibration
Pressure and
Temperature
Measurement
(Version 2.0
and higher)
aM7!
aD0!
attt2
ttt is the time in seconds until the
measurement is ready and 2 is the
number of values that can be collected
a
service request
apt
where p is the pressure psi and t is the
temperature in degrees Celsius.
Bringing the Benefits of Real-Time Data Collection to the World
Sutron Corporation, 21300 Ridgetop Circle, Sterling, Virginia 20166-6520
31
Содержание Accububble 56-0131-50-1
Страница 2: ......