Note that the command follows the SDI-12 standard beginning with the address and ending with
"!".
The ACCUBUBBLE will issue a reply message in response to the command if the command was
recognized. The message will be 00011 which is explained in the Command Reference. If you do
not get this message, try the command again and check the switches (Unit must be set to address 0
since that is the address this command trying to change from). Note: The ACCUBUBBLE will not
respond if the command is invalid, i.e., there is a typing mistake in the command or the two copies
of the new address do not match.
As an example, the following command would set the ACCUBUBBLE address to 5:
0XAD5A5!
Subsequently, the address can be set to a different address, 9 for example, by the command:
5XAD9A9!
Beginning with version 1.9 the ACCUBUBBLE also supports an alternate version of the set
Address command as specified in SDI-12 standard version 1.2.
0An!
where 0 is the current address of the device, n is the new SDI-12 address
(0 to 9, A to Z, a to z).
As an example, the following command would set the ACCUBUBBLE address to 5:
0A5!
The ACCUBUBBLE will respond with the new address which is 5.
Subsequently, the address can be set to a different address, 9 for example, by the command:
5A9!
Verifying the Address and Operation
The ACCUBUBBLE will respond with an identifying message when it receives the
send
identification
command, I. The format of the command is:
aI!
Where a is the address for the ACCUBUBBLE.
The ACCUBUBBLE will reply with
a13 SUTRON 0131-31.0sssssssVvvv
a
13
SUTRON
0131-3
1.0
sssssss
Vvvv
Where:
SDI-12 address
supports SDI version 1.3 commands
manufacturer SUTRON
Sutron model number
hardware revision level
sensor serial number
the software revision