Quadrature Data Recorder Connection (Shaft Encoder Replacement)
5600-0131–3 and 56-0131-50-1 only
Description Terminal
Block
Notes
Pump Power IN Positive
1
+8 to +16 VDC (3 Amp)
Pump Power IN Ground
2
Sensor Pwr In +
3
8-28 VDC Power connection to data recorder
Sensor Pwr In -
4
Ground Connection to Data Recorder
SDI-12 Data
5
Analog output (A+)
6
Analog ground (A-)
7
Quadrature out Phase A
8
Phase A will lead Phase B for positive change (only on –3)
Quadrature out Phase B
9
Phase A leads by going to 5V before Phase B (only on –3)
Note: The XQS command should be issued to the unit to correctly configure the quadrature output step size and rate. The
XQC command can be used to synchronize the output with the quadrature recording device. The XOM command
specifies whether the Quadrature output should be updated independently of SDI-12.
Analog Display or Data Recorder Connection
5600-0131-4 and 56-0131-50-2 only
Description Terminal
Block
Notes
Pump Power IN Positive
1
+8 to +16 VDC (3 Amp)
Pump Power IN Ground
2
Sensor Pwr In +
3
8-28 VDC Power connection to data recorder
Sensor Pwr In -
4
Ground Connection to Data Recorder
SDI-12 Data
5
Analog output (A+)
6
0-5 VDC (only on –4)
Analog ground (A-)
7
Output ground reference
Quadrature out Phase A
8
Quadrature out Phase B
9
Note: The XAR command command can be utilized to customize the analog output range. The XOM command specifies
whether the Analog output should be updated independently of SDI-12.