Command
Description
Command Syntax
(command underlined)
Sensor response (underlined)
"a" represents the
single-character address
CC4 Request
Standards lab
Calibration
Scale and
Offset with
CRC-16
(Version 2.0
and higher)
aCC4!
aD0!
a00002
000 is the time in seconds until the
measurement is ready and 2 is the
number of values that can be collected
asoC
where, s is the scale calibration, o is the
offset calibration(psi), and C is the CRC.
CC5 Request
Quadrature
scale factor,
threshold,
step rate, and
operating
mode for the
unit with
CRC-16
(Version 2.0
and higher)
aCC5!
aD0!
a0004
000 is the time in seconds until the
measurement is ready and 4 is the
number of values that can be collected
astrmC
where, s is the Quadrature scale, t is the
quadrature threshold, r is the quadrature
step rate in steps per second, and m is
the operating mode of the analog and
quadrature outputs, and C is the CRC.
CC6 Request
Concurrent
Temperature
and Pressure
Measurement
with CRC-16
(Version 2.0
and higher)
aCC6!
aD0!
attt04
ttt is the time in seconds until the
measurement is ready and 04 is the
number of values that can be collected
atupvC
where t is the temperature, u is the
temperature units, p is the pressure, v is
the pressure units, and C is the CRC.
Use the XUT command to set the
temperature units and the XUP
command to set the pressure units.
CC7 Request
Concurrent
factory
calibration
Pressure and
Temperature
Measurement
with CRC-16
(Version 2.0
and higher)
aCC7!
aD0!
attt02
ttt is the time in seconds until the
measurement is ready and 02 is the
number of values that can be collected
aptC
where p is the pressure psi, t is the
temperature in degrees Celsius, and C is
the CRC.