19
★
Timing printing: Set the time for timing printing .
★
Starting time: Set the starting time of timing printing.
★
Alarming printing: On and Off are optional.
★
Clear data: Clear all stored data in instrument memory, including historical data, alarming
list, file list, power-off record.
Operation: in this screen,
Move the cursor according to the "Left shift" and "Right shift" keys.
Adjust the numeric value according to "Down shift", "Up shift", " + " and "-".
Press the "Enter" key to confirm the execution; when the cursor moves to clear the data, press
the " Enter" key and pop up the dialog box “whether to clear all data", select "Yes" to perform the
clearance of historical data, and choose "No" to abandon the clearance of historical data.
Press the "Esc" key to exit this screen. If the content of the screen is modified, it will pop up
the dialogue box "Do you save the modified parameters?" In the dialog box, select Yes to save the
modified parameters and No to save the modified parameters.
4.12.2 Record configuration
★
Record mode: non-circulation mode and circulation.
Non-circulation: Automatically stop recording when the recording space of the instrument is
full or reach 100 pieces.
Circulation: When the record space is full, it will continue to record from the first
file .According to the way of replacing the old file with the new file, when the number of records
reaches 100, the first file will be deleted, the serial number of the other files will move forward in
turn, and then create a new file to continue recording.
★
Recording interval: The interval such as 1 second, 2 seconds, 4seconds, 6 seconds, 15
seconds, 30 seconds, 1 minute, 2 minutes, 4 minutes is optional.
★
Recording types: Real-time value, average value, maximum value and minimum value are
optional.
Real-time value: Directly take the real-time value corresponding to the time of recording
point.