User Guide
Страница 1: ...User Guide ...
Страница 2: ...User s Manual SunTech 247 Blood Pressure Device for Automated and Manual Measurement ...
Страница 3: ... device with SpO2 SunTech 247 BP Temperature SpO2 BP device with SpO2 and temperature SunTech 247 Battery BP BP device with rechargeable battery SunTech 247 Battery BP Temperature BP device with temperature and rechargeable battery SunTech 247 Battery BP SpO2 BP device with SpO2 and rechargeable battery SunTech 247 Battery BP Temperature SpO2 BP device with SpO2 temperature and rechargeable batter...
Страница 4: ...nual is the proprietary information of SunTech Medical and is provided solely for purposes of operation maintenance or service of the SunTech 247 This manual and the SunTech 247 described in it are protected under copyright law under which they may not be copied in whole or in part without written consent of SunTech Medical SunTech is a registered trademark of SunTech Medical Inc All other tradema...
Страница 5: ...ood pressure EMR EHR Electronic Medical Records Electronic Health Records Document Key This manual uses the following icons to call attention to specific instructions or guidance TIP A step or process that eases or enhances your use of your SunTech 247 device NOTE Indicates something you must do to use your device correctly and effectively CAUTION Warns you that not following these instructions ca...
Страница 6: ...er knowingly use a defective device Immediately replace parts that are broken worn missing incomplete damaged or contaminated Contact the nearest factory approved service center should repair or replacement become necessary A list of approved service centers appears on page 46 or on our website at www SunTechMed com Further the user of the device bears sole responsibility for any malfunction that ...
Страница 7: ... list of recommended accessories is on page 61 Do not operate the SunTech 247 near flammable anesthetics or volatile vapors An explosion may result Compressing the pneumatic tubing may cause system errors Do not use the device if it has failed its diagnostic self test or if it displays a greater than zero pressure with no cuff attached or a value of functional oxygen saturation or temperature with...
Страница 8: ...ing it back into use The SunTech 247 is not intended for patients connected to a cardiopulmonary bypass machine At least every three months inspect probes cords and accessories for fraying or other mechanical damage Replace as necessary Check the calibration of your SunTech 247 at least once every two years If Luer Lock connectors are used in the constructions of tubing there is a possibility that...
Страница 9: ...s 2 Other Accessories 2 A Bird s Eye View 3 BP module 3 BP display 5 Optional temperature and pulse oximetry modules 6 Optional temperature and pulse oximetry displays 8 Icons and Cues 9 Auditory Cues 9 Visual Cues Battery Icon 10 Visual Cues Blood Pressure Module 10 Visual Cues Temperature Module 11 Visual Cues Pulse Oximetry Module 12 QUICK START GUIDE 14 Measuring Blood Pressure and Pulse Rate ...
Страница 10: ...SUNTECH 247 20 Prepping Your Patient 20 Selecting the Right Cuff 21 Taking a Measurement 22 Taking an Automated Measurement 22 Taking a Sphygmode Measurement 23 K Sounds A Primer 24 MEASURING TEMPERATURE WITH THE SUNTECH 247 25 Temperature Units of Measurement 25 Temperature Measurement Modes 25 Using temperature probe and probe covers 25 Taking an Oral Temperature 26 Taking an Axillary Temperatur...
Страница 11: ...HOOTING 40 Troubleshooting Blood Pressure Module 40 Troubleshooting Temperature Module 41 Troubleshooting Pulse Oximetry Module 42 FAQs 44 Web Resources 46 Service Centers 46 For customers in the Americas 46 For customers in Europe the Middle East and Africa 47 For customers in Asia and the Pacific 47 SPECIAL SITUATIONS 48 Special Situations 48 Measuring Blood Pressure in Children 48 Measuring Blo...
Страница 12: ...xi APPENDICES 50 Specifications 50 Compliance 51 Safety Requirements 51 EMC Statement 52 Symbols 52 EU and FCC Regulatory Information 58 Limited Warranty 59 Purchasing Parts and Accessories 61 ...
Страница 13: ...low Upon opening your kit please ensure that all listed contents are included If any contents are missing or damaged please contact SunTech The SunTech 247 Kit The SunTech 247 kit contains your AC powered device Your kit will also contain An 8 foot blood pressure hose Adult and large adult size all purpose cuffs A wall mounting kit A power supply A geography specific power cord A CD with this manu...
Страница 14: ...lcor compatible pulse oximetry module with an adult reusable finger sensor 3ft 0 9m length and sensor extension cable 6ft 1 82m length A vSat pulse oximetry module with an adult reusable finger sensor 10ft 3m length Other Accessories Many other accessories and sources for purchasing them are listed in the Appendix on page 61 A few to note include An APC Adult package contains one each of the follo...
Страница 15: ...er supply Blood Pressure Hose Connector Connects to the 8 foot pressure hose Buttons The buttons are used for all actions on the interface and provide tactile feedback when pressed Automatic button SphygMode manual button Power button recessed Blood pressure hose connector Power connector rear panel ...
Страница 16: ...dy to take measurements SphygMode Manual Idle Select Inflates the cuff as long as the button is selected Taking a SphygMode BP Select Re inflates the cuff as long as the button is selected Taking an automatic BP Select Aborts the BP measurement in progress In Calibration Check mode Select Device exits Calibration Check mode and is ready to take measurements Power Power off Select Turns on the devi...
Страница 17: ...ulse rate If the cuff hose error and warning icons are lit a status code may appear in this space See page 41 for details Pulse rate icon When displayed the value below is the patient s pulse rate Mean arterial pressure icon When displayed the value below is the patient s MAP Power level indicator Displays connection to AC power or for the battery version the charge level of the battery Cuff hose ...
Страница 18: ...imetry modules Buttons on the optional temperature module Temperature units select Recessed momentary switch that toggles the display between F and C Temperature mode select symbol on button Momentary switch that selects oral vs axillary measurements for the oral temperature probe and selects predictive vs direct measurement methods for all probes Temperature units select recessed Temperature mode...
Страница 19: ...nects to an SpO2 extension cable or sensor See page 62 for compatible accessories All vSat SpO2 modules can be identified by a blue label around the SpO2 sensor connector Only vSat SpO2 sensors blue sensor cable and connector should be used with vSat modules All Nellcor compatible SpO2 modules can be identified by a white label around the SpO2 sensor connector Only Nellcor compatible sensors shoul...
Страница 20: ...ype and related information If the warning icon is lit displays the status code See page 41 for details Celsius icon When lit the value below is displayed in degrees Celsius Fahrenheit icon When lit the value is displayed in degrees Fahrenheit Warning icon When displayed indicates that the system needs to be checked See page 41 for details Celsius icon Fahrenheit icon Temperature Signal quality Sp...
Страница 21: ...tabulated in this chapter Audible cues or beeps identify stages in the measurement cycle Icons illuminated within a module s display indicate measurement modes processes or warnings Auditory Cues The temperature and BP modules of the SunTech 247 are programmed with auditory cues A listing of these cues appears below NUMBER OF BEEPS INDICATES One short beep after power up or right before powering d...
Страница 22: ...arge level drops the segments will be turned off in sequence from the right to the left The battery charge is very low Recharge before using segment flashing Visual Cues Blood Pressure Module Icons and numeric displays on your device assist you in taking quick and accurate readings ICON DISPLAY INDICATES The systolic BP read in mmHg displays immediately below this symbol The diastolic BP read in m...
Страница 23: ...evice is unable to take a valid reading See page 40 for troubleshooting details A measurement is in progress If the column is rising the cuff is being inflated if the column is falling the cuff is deflating Each segment lit is approximately equivalent to 10 mmHg Visual Cues Temperature Module ICON DISPLAY INDICATES Temperature shown in degrees Fahrenheit Temperature shown in degrees Celsius Travel...
Страница 24: ...t measurement mode but the reading is currently out of range Temperature value flashes in an upward direction Final measurement is greater than 109 4 F 43 0 C Temperature value flashes in a downward direction Final measurement is less than 86 F 30 0 C Temperature value is steady no flashing This is the final temperature value Visual Cues Pulse Oximetry Module ICON DISPLAY INDICATES Traveling dash ...
Страница 25: ...nd quality from the pulse oximeter sensor If there is no measurement and the signal quality is low try a different site or sensor SpO2 value Indicates the functional oxygen saturation This area also displays the status code when the warning symbol is lit ...
Страница 26: ...ep to indicate cycle completion The systolic and diastolic values are shown in their respective locations If MAP mode is enabled the pulse rate and MAP values will alternate Measuring Blood Pressure Using SphygMode 1 Wrap an appropriately sized cuff sizes are tabulated on page 21 snugly around the upper arm midway between the elbow and shoulder 2 Ask the patient to stay still and quiet 3 Press and...
Страница 27: ... display 3 Remove the probe from the patient s mouth discard the probe cover by pressing the button on the end of the probe handle and replace the probe in its holder ready for the next measurement Measuring Oxygen Saturation 1 For the reusable finger sensor insert the patient s digit index most preferable into the sensor You will see a traveling dash until a valid reading is available typically i...
Страница 28: ...lternate device After taking confirmatory readings check the device for proper functioning Arrange the power supply and cabling so that it does not constitute a hazard to your patient your co workers or yourself Protecting Yourself Removing the cover or the back of the device can cause electric shocks Do not attempt to service your SunTech 247 unless you are authorized Protecting Your SunTech 247 ...
Страница 29: ...Attach the wall bracket adapter to the rear panel of the BP device using the 3 screws 3 Position the adapter onto the rivets of the wall bracket and slide the device down until it locks into place The weight of the contents of the wall mounted storage basket should never exceed five pounds Please do not store heavy items in the storage basket Affixing the Device to a Mobile Stand Attaching the Sun...
Страница 30: ... per charge Connecting Your Device To maintain the easy readability and streamlined facade of your SunTech 247 all connections are made through the back or sides of the enclosure Connectors on the BP enclosure are for Blood pressure hose Power supply Connectors on the optional modules are for A pulse oximetry sensor on the pulse oximetry module A temperature probe on the temperature module To conn...
Страница 31: ... device will turn on automatically Powering Up 1 Depress the power button located on the right side of the main enclosure The power up sequence begins All display segments light up for three seconds A short beep indicates that the SunTech 247 is ready 2 Check the status of the power level indicator If the power level indicator shows one segment flashing connect the device to the power supply befor...
Страница 32: ...stricting clothing on the selected arm Has no injury or tissue damage on the selected arm Keeps the cuffed arm at heart level Keeps the cuffed arm motion free and relaxed without any muscle tension in the biceps and triceps during the measurement Does not cross his her legs for the measurement Keep aware of current practices as recommended by the American Heart Association British Hypertension Soc...
Страница 33: ...rence falls within the specified range of the cuff If the arm is within range this cuff size is correct for your patient If the measurement is outside the RANGE indicator use the appropriate larger or smaller cuff and re check Using a cuff that is too small commonly called undercuffing can result in overestimating a patient s BP Using a cuff that is too large or overcuffing can result in underesti...
Страница 34: ...he cuffed arm icon The cuff inflates to approximately 160 mmHg as indicated in the systolic area of the display 2 Once the cuff pressure reaches its target the device controls the deflation and in some instances re inflation of the cuff in order to accurately measure BP The cuff pressure displays in the systolic area and is also indicated by the vertical LED bar to the left When you hear a single ...
Страница 35: ...ure 3 Once the cuff has been inflated to the desired level release the SphygMode manual button The cuff begins to deflate at approximately 3mmHg sec and the device displays the cuff pressure 4 While listening to your stethoscope note the systolic and diastolic pressures corresponding to the first and last Korotkoff sounds K sounds heard Press the SphygMode manual button to re inflate the cuff Pres...
Страница 36: ...imilar to those heard during phase 1 An abrupt muffling of sound identifies K 4 the fourth phase The end or fifth phase is the point at which sounds cease to be heard altogether Systolic pressure is registered at K 1 and diastolic at K 5 K 4 or K 5 There exists some debate about whether K 4 or K 5 should be recorded as the diastolic BP In most cases K 5 is preferred However if the sound persists e...
Страница 37: ...rature Measurement Modes The device can measure temperature via three modes Oral indicated by on the display and measured using the blue probe Axillary indicated by on the display and measured using the blue probe Rectal indicated by on the display and measured using the red probe All three modes can be used for both predictive and direct measurement In the default predictive mode your SunTech 247...
Страница 38: ...obe carefully under the patient s tongue as denoted by the heat pockets shown to the right The posterior medial sublingual pocket is preferred for accuracy Hold the probe in place so that its tip maintains tissue contact Close the patient s mouth During measurement a traveling dash is displayed 3 After 10 15 seconds a long beep signals the end of the measurement cycle The temperature will display ...
Страница 39: ...rm by the patient s side During measurement a traveling dash is displayed Note that proper placement of the probe is essential to the accuracy of the measurement 4 After 10 15 seconds a long beep signals the end of the measurement cycle The temperature will display for two minutes or until you initiate a new measurement 5 Remove the probe discard the probe cover and place back in the probe holder ...
Страница 40: ... traveling dash is displayed Note that proper placement of the probe is essential to the accuracy of the measurement 6 After 10 15 seconds a long beep signals the end of the measurement The result will be displayed for two minutes or until you initiate a new measurement 7 Remove the probe discard the probe cover and place back in the probe holder Note the reading If the temperature reading is out ...
Страница 41: ... including Patient s body weight Patient activity Infection control concerns For most patients greater than 30kg use an adult sensor for patients 10 50kg a pediatric sensor may provide better fit Disposable sensors can provide a more secure connection to the patient Protecting Your Pulse Oximetry Sensors In addition to the safety instructions for your SunTech 247 here are some additional tips on c...
Страница 42: ... medical status or skin condition Discontinue use of adhesive tape strips if the patient exhibits an allergic reaction to the adhesive materials Measuring Oxygen Saturation with Sensor on Finger For best results clip the sensor on the index middle or ring finger avoiding the little finger or thumb 1 For the reusable finger sensor insert the patient s digit index most preferable into the sensor Ens...
Страница 43: ...pO2 determination begins a traveling dash will be displayed until a measurement is determined usually in 10 seconds This reading will be updated once per second SpO2 can be measured without interruption for up to 10 minutes Along with the functional oxygen saturation value the signal strength will also display 2 Note the patient s reading and remove the clip gently When you remove the sensors from...
Страница 44: ...ing recorded at the end of the ten minute period will be the recalled reading Clearing the Last Set of Readings To clear the values from the last automatic BP measurement and the accessories press and hold the automatic button more than 5 seconds Previous values will be displayed momentarily Then the display blanks On redisplay you will see dashes for all the values that have been cleared Your Sun...
Страница 45: ...your equipment Cleaning Cleaning the Device 1 Wipe the device with a soft damp cloth to remove surface dust and dirt The SunTech 247 device cannot be sterilized Never immerse the device in any fluid or attempt to employ cleaning fluids or solvents Cleaning the Cuffs 1 Between uses wipe cuff sleeves and the insides of cuffs with a medical grade cleaning agent 2 Periodically remove the bladders and ...
Страница 46: ...h as a manometer and inflation bulb to the pressure hose connector of the SunTech 247 Contact SunTech Medical customer support for details on ordering the calibration kit which includes a T connector 3 Compare the pressure reference to the SunTech 247 throughout the pressure range 0 to 270mmHg If the difference between the pressure reference and the SunTech 247 is no larger than 2mmHg the SunTech ...
Страница 47: ... pull on the wires attached to the battery terminals 3 Disconnect the wires from the battery terminals 4 The rechargeable battery contains lead Please dispose of the old battery properly 5 Connect the wires to the terminals of the replacement battery ensuring the red wire is attached to the red terminal and the black wire to the black terminal If the wires are reversed no damage will occur however...
Страница 48: ...h the display Part B the probe holder which holds the temperature probe and probe cover box 1 Using the power button on the right side of the BP module ensure that the SunTech 247 is off 2 Remove the cover plate from the top of the BP module 3 Slide part A of the temperature module along the guides on the top of the BP module from front to back until it snaps into place All the segments of the tem...
Страница 49: ...odule At the end of the start up sequence the temperature module display will blank except for the appropriate temperature unit icon F or C A short beep indicates that the Sun Tech 247 is ready Note If the temperature module does not appear to be working properly cycle the power several times using the power button on the right side of the BP module This will synchronize all the modules The module...
Страница 50: ...d of the start up sequence the pulse oximetry module display will blank except for the SpO2 symbol 5 Turn the device off Connect the sensor or extension cable to the connector on the module Secure the cable to your device using the retention clip 6 Turn the device on All display segments light up for three to five seconds A short beep indicates that the SunTech 247 is ready Note If the pulse oxime...
Страница 51: ...he device Moving Your Device To pack your device for repair or transport 1 Detach the cuff temperature probe SpO2 sensor power supply and other ancillary products from the device 2 Disconnect the battery and remove it from the device 3 Place the device in the original shipping carton preferably with its original packing material 4 Ensure that the device will be kept at between 20 C 4 F and 55 C 13...
Страница 52: ...less system While the Bluetooth capability on the 247 has been rated up to a maximum range of 10m be sure to minimize walls structures and other obstacles that may impede wireless connectivity Troubleshooting Blood Pressure Module Problem Wrong size cuff Misplaced cuff or Blocked brachial artery Solutions 1 Check that the cuff is in the correct position 2 Check that the cuff is properly tightened ...
Страница 53: ...er down then power up the unit using the power button on the right side of the unit If the error does not recur immediately take another measurement 3 If the error recurs remove power from the unit unplug any power supplies and remove the battery if there is one then reconnect the power If the error does not recur immediately take another measurement 4 If the error recurs contact SunTech Medical I...
Страница 54: ...t button and or the Temperature mode select button until the button becomes unstuck When the button is unstuck the error will no longer be displayed If you cannot un stick the button contact SunTech or an authorized service center Status Code 20 Problem Hardware error Solutions 1 Take another measurement 2 If this is unsuccessful power down then power up the unit using the power button on the righ...
Страница 55: ...m Defective sensor Solution Replace the sensor and take a new measurement Status Code 20 Problem Hardware error Solutions 1 Take another measurement 2 If this is unsuccessful power down then power up the unit using the power button on the right side of the unit If the error does not recur immediately take another measurement 3 If the error recurs remove power from the unit unplug any power supplie...
Страница 56: ... of polarities Replace the cover securely and connect the device to AC mains power via the power supply to turn the device on Ensure the battery is fully charged before use See page 35 for detailed instructions How often should I calibrate the BP Module for the SunTech 247 You should check the calibration every two years If there is a difference larger than 2 mmHg against the pressure reference th...
Страница 57: ...Should I wait between temperature measurements Yes Accurate temperature measurement requires the probe to be at normal room temperature After taking a measurement wait for the probe to return to room temperature or wipe the probe with an alcohol wipe before taking a subsequent measurement Do I need to calibrate the temperature or pulse oximetry modules For calibration or service on the SunTech 247...
Страница 58: ... Bluetooth capability to allow communication with any EMR EHR system Please contact your IT administrator or EMR EHR system provider about configuring your SunTech 247 for communication Software development kits are available from SunTech Medical for EMR EHR vendors to support communication with the SunTech 247 Web Resources www SunTechMed com Service Centers For customers in the Americas SunTech ...
Страница 59: ...ical Ltd Oakfield Estate Eynsham Oxfordshire England Tel 44 0 1865 884 234 Fax 44 0 1865 884 235 For customers in Asia and the Pacific SunTech Medical Ltd Level 19 Two International Finance Centre 8 Finance Street Central Hong Kong Tel 852 2251 1949 Fax 852 2251 1950 ...
Страница 60: ...re in Obese Patients There appears to be a positive correlation between obesity and hypertension Due to the increased arm circumference of obese patients use of a standard cuff may lead to blood pressure being erroneously elevated a condition known as cuff hypertension Selecting an Appropriate Cuff for Obese Patients For larger than normal upper arms use a wider and longer cuff than you would othe...
Страница 61: ...s Measuring Blood Pressure in the Emergency Room Measuring blood pressure in the emergency room can be done through automated blood pressure measurements For critically ill or injured patients blood pressure should be measured through the invasive arterial pressure method Measuring Blood Pressure in the Presence of Orthostatic Hypotension Orthostatic hypotension is defined as a decrease in systoli...
Страница 62: ...bpm 2 or 3 bpm whichever is greater Temperature range 86 F 30 0 C 109 4 F 43 0 C Temperate accuracy 0 2 F 0 1 C Functional oxygen saturation range 40 100 Functional oxygen saturation accuracy 70 100 2 digits Note because pulse oximeter equipment measurements are statistically distributed only about two thirds of pulse oximeter equipment measurements can be expected to fall within 2 digits of the v...
Страница 63: ...I SP10 2002 ASTM E 1112 EN 12470 3 Meets EN 1060 1 Specification for non invasive sphygmomanometers Part 1 General requirements and EN 1060 3 Non invasive Sphygmomanometers Part 3 Supplementary Requirements for Electro Mechanical BP Measuring Systems Classification Protection against electric shock Class II for non battery version Internally Powered Equipment for battery version Applied parts Type...
Страница 64: ...ar installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving device Increase the separation between the equipment Connect the equipment into an outlet on a circuit different from that ...
Страница 65: ...nying documents Class II isolation equipment or CE approval Recognized component certified by UL to both Canada and US requirements TUV Canada and US approval TUV International approval Earth ground Output connection configuration positive voltage negative shield ...
Страница 66: ... Group 1 The SunTech 247 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The SunTech 247 is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies b...
Страница 67: ...d be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for...
Страница 68: ...itters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection fro...
Страница 69: ...mitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 3 5 V1 P 80 MHz to 800 MHz d 3 5 V1 P 800MHz to 2 5GHz d 7 E1 P 0 01 0 12 0 12 0 23 0 10 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to...
Страница 70: ...one of more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Operation is subject to the following two conditions 1 this device may not cause harmful ...
Страница 71: ...tes prepaid to the factory SunTech Medical Inc will repair any component s or part s that it finds to be defective during the period of this limited warranty Should a defect become apparent the original purchaser should first notify SunTech Medical Inc of the suspected defect The instrument should be carefully packaged and shipped prepaid to For customers in the Americas SunTech Medical Inc Servic...
Страница 72: ...amaged by accident misuse negligence act of God or serviced by any person not authorized by SunTech Medical Inc This limited warranty contains the entire obligation of SunTech Medical Inc and no other warranties expressed implied or statutory are given No representative or employee of SunTech Medical Inc is authorized to assume any further liability or grant any further warranties except as herein...
Страница 73: ...ect connectors 98 0080 02 All Purpose Cuff Child Blood Pressure Cuff with quick connect connector Range 12 19cm Color Green 98 0080 03 All Purpose Cuff Child Long Blood Pressure Cuff with quick connect connector Range 12 19cm Color Green 98 0080 04 All Purpose Cuff Small Adult Blood Pressure Cuff with quick connect connector Range 17 25cm Color Royal Blue 98 0080 05 All Purpose Cuff Small Adult Lo...
Страница 74: ...y sensor Nellcor compatible Reusable finger sensor with 10ft 3 0m cable and DB 9M connector for persons 88 lbs 40kg 52 0005 01 Y multi site reusable oximetry sensor Nellcor compatible Reusable universal Y sensor including ear clip with 36 91 m cable and DB 9M connector for all patients 2 2lbs 1 kg 52 0005 04 6 foot SpO2 Sensor Extension Cable Nellcor compatible Reusable cables with DB 9 M F connec...
Страница 75: ...e Rectal probe with 9 ft 2 7 m extended cord 98 0147 00 Red Rectal Well Rear piece of the temperature module that holds the rectal probe and box of probe covers 98 0130 01 Disposable Probe Covers 5000 10 trays 25 boxes tray 20 probe covers box 98 0130 01 Disposable Probe Covers 5000 250 boxes 20 probe covers box 98 0131 01 Disposable Probe Covers 500 25 boxes 20 probe covers box 52 0004 02 Calibra...
Страница 76: ... 60 Hz Output 9 V medical grade 91 0003 05 EU power cord Power or mains lead with Type E and F hybrid CEE 7 7 two pin plug 8 2 ft 2 5 m length 91 0003 06 UK power cord Power or mains lead with Type G BS1363 three pin plug 8 2 ft 2 5 m length 91 0003 00 US power cord Power or mains lead with Type B NEMA 5 15 three pin plug 8 2 ft 2 5m length 91 0097 00 BP hose BP hose with mate for threaded screw t...
Страница 77: ...00 421 8626 Fax 1 919 654 2301 SunTech Medical Ltd Level 19 Two International Finance Centre 8 Finance Street Central Hong Kong Tel 852 2251 1949 Fax 852 2251 1950 SunTech Medical Ltd Service Department Oakfield Industrial Estate Eynsham Oxfordshire OX29 4TS UK Tel 44 1865 884 234 Fax 44 1865 884 235 Part 80 0040 00 Rev G ...
Страница 78: ...or Puritan Bennett LLC FILAC and FASTEMP are registered trademarks of Tyco Healthcare Group LP The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by SunTech Medical Inc is under license ...